BAHA devices and magnetic resonance imaging scanners

Michael H. Fritsch, Ilka C. Naumann, Kristine M. Mosier

Research output: Contribution to journalReview article

9 Citations (Scopus)

Abstract

Hypothesis: That an osteointegrated Bone-anchored Hearing Aid (BAHA System) will show movement or displacement when exposed to a 9.4-T magnetic resonance imaging (MRI) magnet and that any movement may have implications for patient safety and image quality. Background: BAHA System implanted patients may require MRI scans for various pathologies. The possibility of BAHA fixture and abutment movement can be explored to help clarify patients' and physicians' decisions regarding safety issues. Setting: University Medical Center (tertiary care center). Methods: BAHA System assembled fixture with abutment of 3 and 4 mm were each placed inside of a Petri dish and mounted into human cadaver temporal bones, then exposed to a 9.4-T MRI magnet. The BAHA System external processor was entered into the MRI suite. Head MRI images and charts of 2 patients with BAHA System devices who had undergone MRI scanning at 1.5 and 3.0 T, respectively, were reviewed. Results: The BAHA System fixture with abutment moved inside of the Petri dish. The temporal bone studies revealed no sign of motion or displacement. The external BAHA processor showed strong movement toward the magnet. MRI scannings of patients were without any adverse effects. Image MRI quality was affected only around the BAHA post. Conclusion: The bone-mounted fixture with abutment of 3 and 4 mm resisted any magnetic forces trying to displace them up to 9.4 T. Two patients underwent scanning at 1.5 and 3.0 T with no adverse effects. MRI image quality was excellent starting at 1 cm from the fixture with abutment. The study findings lend strong evidence that patients can be safely scanned up to 9.4 T without adverse effects.

Original languageEnglish (US)
Pages (from-to)1095-1099
Number of pages5
JournalOtology and Neurotology
Volume29
Issue number8
DOIs
StatePublished - Dec 1 2008

Fingerprint

Magnetic Resonance Imaging
Equipment and Supplies
Magnets
Temporal Bone
Bone and Bones
Hearing Aids
Patient Safety
Cadaver
Tertiary Care Centers
Head
Pathology
Physicians
Safety

Keywords

  • Baha
  • Hearing aid
  • Implant
  • Magnetic resonance imaging
  • Safety

ASJC Scopus subject areas

  • Otorhinolaryngology
  • Clinical Neurology
  • Sensory Systems
  • Medicine(all)

Cite this

BAHA devices and magnetic resonance imaging scanners. / Fritsch, Michael H.; Naumann, Ilka C.; Mosier, Kristine M.

In: Otology and Neurotology, Vol. 29, No. 8, 01.12.2008, p. 1095-1099.

Research output: Contribution to journalReview article

Fritsch, Michael H. ; Naumann, Ilka C. ; Mosier, Kristine M. / BAHA devices and magnetic resonance imaging scanners. In: Otology and Neurotology. 2008 ; Vol. 29, No. 8. pp. 1095-1099.
@article{988489e4ae964323b0005a011d8c83de,
title = "BAHA devices and magnetic resonance imaging scanners",
abstract = "Hypothesis: That an osteointegrated Bone-anchored Hearing Aid (BAHA System) will show movement or displacement when exposed to a 9.4-T magnetic resonance imaging (MRI) magnet and that any movement may have implications for patient safety and image quality. Background: BAHA System implanted patients may require MRI scans for various pathologies. The possibility of BAHA fixture and abutment movement can be explored to help clarify patients' and physicians' decisions regarding safety issues. Setting: University Medical Center (tertiary care center). Methods: BAHA System assembled fixture with abutment of 3 and 4 mm were each placed inside of a Petri dish and mounted into human cadaver temporal bones, then exposed to a 9.4-T MRI magnet. The BAHA System external processor was entered into the MRI suite. Head MRI images and charts of 2 patients with BAHA System devices who had undergone MRI scanning at 1.5 and 3.0 T, respectively, were reviewed. Results: The BAHA System fixture with abutment moved inside of the Petri dish. The temporal bone studies revealed no sign of motion or displacement. The external BAHA processor showed strong movement toward the magnet. MRI scannings of patients were without any adverse effects. Image MRI quality was affected only around the BAHA post. Conclusion: The bone-mounted fixture with abutment of 3 and 4 mm resisted any magnetic forces trying to displace them up to 9.4 T. Two patients underwent scanning at 1.5 and 3.0 T with no adverse effects. MRI image quality was excellent starting at 1 cm from the fixture with abutment. The study findings lend strong evidence that patients can be safely scanned up to 9.4 T without adverse effects.",
keywords = "Baha, Hearing aid, Implant, Magnetic resonance imaging, Safety",
author = "Fritsch, {Michael H.} and Naumann, {Ilka C.} and Mosier, {Kristine M.}",
year = "2008",
month = "12",
day = "1",
doi = "10.1097/MAO.0b013e31818201fd",
language = "English (US)",
volume = "29",
pages = "1095--1099",
journal = "Otology and Neurotology",
issn = "1531-7129",
publisher = "Lippincott Williams and Wilkins",
number = "8",

}

TY - JOUR

T1 - BAHA devices and magnetic resonance imaging scanners

AU - Fritsch, Michael H.

AU - Naumann, Ilka C.

AU - Mosier, Kristine M.

PY - 2008/12/1

Y1 - 2008/12/1

N2 - Hypothesis: That an osteointegrated Bone-anchored Hearing Aid (BAHA System) will show movement or displacement when exposed to a 9.4-T magnetic resonance imaging (MRI) magnet and that any movement may have implications for patient safety and image quality. Background: BAHA System implanted patients may require MRI scans for various pathologies. The possibility of BAHA fixture and abutment movement can be explored to help clarify patients' and physicians' decisions regarding safety issues. Setting: University Medical Center (tertiary care center). Methods: BAHA System assembled fixture with abutment of 3 and 4 mm were each placed inside of a Petri dish and mounted into human cadaver temporal bones, then exposed to a 9.4-T MRI magnet. The BAHA System external processor was entered into the MRI suite. Head MRI images and charts of 2 patients with BAHA System devices who had undergone MRI scanning at 1.5 and 3.0 T, respectively, were reviewed. Results: The BAHA System fixture with abutment moved inside of the Petri dish. The temporal bone studies revealed no sign of motion or displacement. The external BAHA processor showed strong movement toward the magnet. MRI scannings of patients were without any adverse effects. Image MRI quality was affected only around the BAHA post. Conclusion: The bone-mounted fixture with abutment of 3 and 4 mm resisted any magnetic forces trying to displace them up to 9.4 T. Two patients underwent scanning at 1.5 and 3.0 T with no adverse effects. MRI image quality was excellent starting at 1 cm from the fixture with abutment. The study findings lend strong evidence that patients can be safely scanned up to 9.4 T without adverse effects.

AB - Hypothesis: That an osteointegrated Bone-anchored Hearing Aid (BAHA System) will show movement or displacement when exposed to a 9.4-T magnetic resonance imaging (MRI) magnet and that any movement may have implications for patient safety and image quality. Background: BAHA System implanted patients may require MRI scans for various pathologies. The possibility of BAHA fixture and abutment movement can be explored to help clarify patients' and physicians' decisions regarding safety issues. Setting: University Medical Center (tertiary care center). Methods: BAHA System assembled fixture with abutment of 3 and 4 mm were each placed inside of a Petri dish and mounted into human cadaver temporal bones, then exposed to a 9.4-T MRI magnet. The BAHA System external processor was entered into the MRI suite. Head MRI images and charts of 2 patients with BAHA System devices who had undergone MRI scanning at 1.5 and 3.0 T, respectively, were reviewed. Results: The BAHA System fixture with abutment moved inside of the Petri dish. The temporal bone studies revealed no sign of motion or displacement. The external BAHA processor showed strong movement toward the magnet. MRI scannings of patients were without any adverse effects. Image MRI quality was affected only around the BAHA post. Conclusion: The bone-mounted fixture with abutment of 3 and 4 mm resisted any magnetic forces trying to displace them up to 9.4 T. Two patients underwent scanning at 1.5 and 3.0 T with no adverse effects. MRI image quality was excellent starting at 1 cm from the fixture with abutment. The study findings lend strong evidence that patients can be safely scanned up to 9.4 T without adverse effects.

KW - Baha

KW - Hearing aid

KW - Implant

KW - Magnetic resonance imaging

KW - Safety

UR - http://www.scopus.com/inward/record.url?scp=59849116697&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=59849116697&partnerID=8YFLogxK

U2 - 10.1097/MAO.0b013e31818201fd

DO - 10.1097/MAO.0b013e31818201fd

M3 - Review article

C2 - 18636027

AN - SCOPUS:59849116697

VL - 29

SP - 1095

EP - 1099

JO - Otology and Neurotology

JF - Otology and Neurotology

SN - 1531-7129

IS - 8

ER -