Abstract

Bone remodeling around cementless acetabular components after total hip arthroplasty has not been well characterized. A randomized, prospective study of total hip arthroplasty was performed comparing 2 cementless acetabular implants: a solid titanium and a more elastic porous tantalum design. Seventeen hips (9 porous tantalum, 8 titanium) underwent quantitative computed tomography at mean of 7.7 years, and adjacent bone mineral density (BMD) was calculated. The absolute and relative decrease in BMD from preoperative level was less in zones 9 to 15 mm adjacent to the porous tantalum compared to the titanium component (P ≤ .02) and predominated posterosuperiorly. The relative BMD increased in all regions adjacent to the porous tantalum component from 5% to 40% over the control. This data demonstrates stress-shielding likely occurs less around a highly porous metal implant of material with an elastic modulus similar to bone.

Original languageEnglish (US)
Pages (from-to)741-747
Number of pages7
JournalJournal of Arthroplasty
Volume25
Issue number5
DOIs
StatePublished - Aug 2010
Externally publishedYes

Fingerprint

Tantalum
Bone Remodeling
Titanium
Metals
Bone Density
Hip
Arthroplasty
Elastic Modulus
Tomography
Prospective Studies
Bone and Bones

Keywords

  • Bone mineral density
  • Bone-remodeling
  • Cementless
  • Porous tantalum
  • Quantitative computed tomography
  • Total hip arthroplasty

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

Bone Remodeling Around Porous Metal Cementless Acetabular Components. / Meneghini, R.; Ford, Kerry S.; McCollough, Cynthia H.; Hanssen, Arlen D.; Lewallen, David G.

In: Journal of Arthroplasty, Vol. 25, No. 5, 08.2010, p. 741-747.

Research output: Contribution to journalArticle

Meneghini, R. ; Ford, Kerry S. ; McCollough, Cynthia H. ; Hanssen, Arlen D. ; Lewallen, David G. / Bone Remodeling Around Porous Metal Cementless Acetabular Components. In: Journal of Arthroplasty. 2010 ; Vol. 25, No. 5. pp. 741-747.
@article{0272ad2eaf84457f94cba64a10a35bff,
title = "Bone Remodeling Around Porous Metal Cementless Acetabular Components",
abstract = "Bone remodeling around cementless acetabular components after total hip arthroplasty has not been well characterized. A randomized, prospective study of total hip arthroplasty was performed comparing 2 cementless acetabular implants: a solid titanium and a more elastic porous tantalum design. Seventeen hips (9 porous tantalum, 8 titanium) underwent quantitative computed tomography at mean of 7.7 years, and adjacent bone mineral density (BMD) was calculated. The absolute and relative decrease in BMD from preoperative level was less in zones 9 to 15 mm adjacent to the porous tantalum compared to the titanium component (P ≤ .02) and predominated posterosuperiorly. The relative BMD increased in all regions adjacent to the porous tantalum component from 5{\%} to 40{\%} over the control. This data demonstrates stress-shielding likely occurs less around a highly porous metal implant of material with an elastic modulus similar to bone.",
keywords = "Bone mineral density, Bone-remodeling, Cementless, Porous tantalum, Quantitative computed tomography, Total hip arthroplasty",
author = "R. Meneghini and Ford, {Kerry S.} and McCollough, {Cynthia H.} and Hanssen, {Arlen D.} and Lewallen, {David G.}",
year = "2010",
month = "8",
doi = "10.1016/j.arth.2009.04.025",
language = "English (US)",
volume = "25",
pages = "741--747",
journal = "Journal of Arthroplasty",
issn = "0883-5403",
publisher = "Churchill Livingstone",
number = "5",

}

TY - JOUR

T1 - Bone Remodeling Around Porous Metal Cementless Acetabular Components

AU - Meneghini, R.

AU - Ford, Kerry S.

AU - McCollough, Cynthia H.

AU - Hanssen, Arlen D.

AU - Lewallen, David G.

PY - 2010/8

Y1 - 2010/8

N2 - Bone remodeling around cementless acetabular components after total hip arthroplasty has not been well characterized. A randomized, prospective study of total hip arthroplasty was performed comparing 2 cementless acetabular implants: a solid titanium and a more elastic porous tantalum design. Seventeen hips (9 porous tantalum, 8 titanium) underwent quantitative computed tomography at mean of 7.7 years, and adjacent bone mineral density (BMD) was calculated. The absolute and relative decrease in BMD from preoperative level was less in zones 9 to 15 mm adjacent to the porous tantalum compared to the titanium component (P ≤ .02) and predominated posterosuperiorly. The relative BMD increased in all regions adjacent to the porous tantalum component from 5% to 40% over the control. This data demonstrates stress-shielding likely occurs less around a highly porous metal implant of material with an elastic modulus similar to bone.

AB - Bone remodeling around cementless acetabular components after total hip arthroplasty has not been well characterized. A randomized, prospective study of total hip arthroplasty was performed comparing 2 cementless acetabular implants: a solid titanium and a more elastic porous tantalum design. Seventeen hips (9 porous tantalum, 8 titanium) underwent quantitative computed tomography at mean of 7.7 years, and adjacent bone mineral density (BMD) was calculated. The absolute and relative decrease in BMD from preoperative level was less in zones 9 to 15 mm adjacent to the porous tantalum compared to the titanium component (P ≤ .02) and predominated posterosuperiorly. The relative BMD increased in all regions adjacent to the porous tantalum component from 5% to 40% over the control. This data demonstrates stress-shielding likely occurs less around a highly porous metal implant of material with an elastic modulus similar to bone.

KW - Bone mineral density

KW - Bone-remodeling

KW - Cementless

KW - Porous tantalum

KW - Quantitative computed tomography

KW - Total hip arthroplasty

UR - http://www.scopus.com/inward/record.url?scp=77954656635&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77954656635&partnerID=8YFLogxK

U2 - 10.1016/j.arth.2009.04.025

DO - 10.1016/j.arth.2009.04.025

M3 - Article

VL - 25

SP - 741

EP - 747

JO - Journal of Arthroplasty

JF - Journal of Arthroplasty

SN - 0883-5403

IS - 5

ER -