Can Intraoperative Sensors Determine the "Target" Ligament Balance? Early Outcomes in Total Knee Arthroplasty

R. Meneghini, Mary M. Ziemba-Davis, Luke R. Lovro, Phillip H. Ireland, Brent M. Damer

Research output: Contribution to journalArticle

42 Scopus citations

Abstract

Background: The optimal "target" ligament balance for each patient undergoing total knee arthroplasty (TKA) remains unknown. The study purpose was to determine if patient outcomes are affected by intraoperative ligament balance measured with force-sensing implant trials and if an optimal "target" balance exists. Methods: A multicenter, retrospective study reviewed consecutive TKAs performed by 3 surgeons. TKA's were performed with standard surgical techniques and ligament releases. After final implants were made, sensor-embedded smart tibial trials were inserted, and compartment forces recorded throughout the range of motion. Clinical outcome measures were obtained preoperatively and at 4 months. Statistical analysis correlated ligament balance with clinical outcomes. Results: One hundred eighty-nine consecutive TKAs were analyzed. Patients were grouped by average medial and lateral compartment force differences. Twenty-nine TKAs (15%) were balanced within 15 lbs and 53 (28%) were "balanced" greater than 75 lbs. Greater improvement in University of California Los Angeles activity level was associated with a mediolateral force difference

Original languageEnglish (US)
JournalJournal of Arthroplasty
DOIs
StateAccepted/In press - Dec 21 2015

Keywords

  • Ligament balance
  • Outcomes
  • Sensor
  • Smart tibial trials
  • Total knee arthroplasty

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Fingerprint Dive into the research topics of 'Can Intraoperative Sensors Determine the "Target" Ligament Balance? Early Outcomes in Total Knee Arthroplasty'. Together they form a unique fingerprint.

  • Cite this