Characterization of a new commercial single crystal diamond detector for photon-and proton-beam dosimetry

Yuichi Akino, Archana Gautam, Len Coutinho, Jan Würfel, Indra J. Das

Research output: Contribution to journalArticle

17 Citations (Scopus)

Abstract

A synthetic single crystal diamond detector (SCDD) is commercially available and is characterized for radiation dosimetry in various radiation beams in this study. The characteristics of the commercial SCDD model 60019 (PTW) with 6- and 15-MV photon beams, and 208-MeV proton beams, were investigated and compared with the pre-characterized detectors: Semiflex (model 31010) and PinPoint (model 31006) ionization chambers (PTW), the EDGE diode detector (Sun Nuclear Corp) and the SFD Stereotactic Dosimetry Diode Detector (IBA). To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s. The depth-dose and profiles data were collected for various field sizes and depths. For all radiation types and field sizes, the depth-dose data of the diamond chamber showed identical curves to those of the ionization chambers. The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region. The temperature dependency was within 0.7% in the range of 4-41°C. A dose of 900 cGy and 1200 cGy was needed to stabilize the chamber to the level within 0.5% and 0.2%, respectively. The PTW type 60019 SCDD detector showed suitable characteristics for radiation dosimetry, for relative dose, depth-dose and profile measurements for a wide range of field sizes. However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements.

Original languageEnglish (US)
Pages (from-to)912-918
Number of pages7
JournalJournal of Radiation Research
Volume56
Issue number6
DOIs
StatePublished - 2015

Fingerprint

Diamond
photon beams
proton beams
Photons
dosimeters
Protons
diamonds
detectors
single crystals
Radiometry
dosage
Radiation
chambers
ionization chambers
Solar System
radiation
profiles
diodes
irradiation
Temperature

Keywords

  • Characterization
  • Detector
  • Dosimetry
  • Microdiamond

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiation
  • Health, Toxicology and Mutagenesis

Cite this

Characterization of a new commercial single crystal diamond detector for photon-and proton-beam dosimetry. / Akino, Yuichi; Gautam, Archana; Coutinho, Len; Würfel, Jan; Das, Indra J.

In: Journal of Radiation Research, Vol. 56, No. 6, 2015, p. 912-918.

Research output: Contribution to journalArticle

Akino, Yuichi ; Gautam, Archana ; Coutinho, Len ; Würfel, Jan ; Das, Indra J. / Characterization of a new commercial single crystal diamond detector for photon-and proton-beam dosimetry. In: Journal of Radiation Research. 2015 ; Vol. 56, No. 6. pp. 912-918.
@article{f192057be5624c4e91482ee5f1553390,
title = "Characterization of a new commercial single crystal diamond detector for photon-and proton-beam dosimetry",
abstract = "A synthetic single crystal diamond detector (SCDD) is commercially available and is characterized for radiation dosimetry in various radiation beams in this study. The characteristics of the commercial SCDD model 60019 (PTW) with 6- and 15-MV photon beams, and 208-MeV proton beams, were investigated and compared with the pre-characterized detectors: Semiflex (model 31010) and PinPoint (model 31006) ionization chambers (PTW), the EDGE diode detector (Sun Nuclear Corp) and the SFD Stereotactic Dosimetry Diode Detector (IBA). To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s. The depth-dose and profiles data were collected for various field sizes and depths. For all radiation types and field sizes, the depth-dose data of the diamond chamber showed identical curves to those of the ionization chambers. The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region. The temperature dependency was within 0.7{\%} in the range of 4-41°C. A dose of 900 cGy and 1200 cGy was needed to stabilize the chamber to the level within 0.5{\%} and 0.2{\%}, respectively. The PTW type 60019 SCDD detector showed suitable characteristics for radiation dosimetry, for relative dose, depth-dose and profile measurements for a wide range of field sizes. However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements.",
keywords = "Characterization, Detector, Dosimetry, Microdiamond",
author = "Yuichi Akino and Archana Gautam and Len Coutinho and Jan W{\"u}rfel and Das, {Indra J.}",
year = "2015",
doi = "10.1093/jrr/rrv044",
language = "English (US)",
volume = "56",
pages = "912--918",
journal = "Journal of Radiation Research",
issn = "0449-3060",
publisher = "Japan Radiation Research Society",
number = "6",

}

TY - JOUR

T1 - Characterization of a new commercial single crystal diamond detector for photon-and proton-beam dosimetry

AU - Akino, Yuichi

AU - Gautam, Archana

AU - Coutinho, Len

AU - Würfel, Jan

AU - Das, Indra J.

PY - 2015

Y1 - 2015

N2 - A synthetic single crystal diamond detector (SCDD) is commercially available and is characterized for radiation dosimetry in various radiation beams in this study. The characteristics of the commercial SCDD model 60019 (PTW) with 6- and 15-MV photon beams, and 208-MeV proton beams, were investigated and compared with the pre-characterized detectors: Semiflex (model 31010) and PinPoint (model 31006) ionization chambers (PTW), the EDGE diode detector (Sun Nuclear Corp) and the SFD Stereotactic Dosimetry Diode Detector (IBA). To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s. The depth-dose and profiles data were collected for various field sizes and depths. For all radiation types and field sizes, the depth-dose data of the diamond chamber showed identical curves to those of the ionization chambers. The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region. The temperature dependency was within 0.7% in the range of 4-41°C. A dose of 900 cGy and 1200 cGy was needed to stabilize the chamber to the level within 0.5% and 0.2%, respectively. The PTW type 60019 SCDD detector showed suitable characteristics for radiation dosimetry, for relative dose, depth-dose and profile measurements for a wide range of field sizes. However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements.

AB - A synthetic single crystal diamond detector (SCDD) is commercially available and is characterized for radiation dosimetry in various radiation beams in this study. The characteristics of the commercial SCDD model 60019 (PTW) with 6- and 15-MV photon beams, and 208-MeV proton beams, were investigated and compared with the pre-characterized detectors: Semiflex (model 31010) and PinPoint (model 31006) ionization chambers (PTW), the EDGE diode detector (Sun Nuclear Corp) and the SFD Stereotactic Dosimetry Diode Detector (IBA). To evaluate the effects of the pre-irradiation, the diamond detector, which had not been irradiated on the day, was set up in the water tank, and the response to 100 MU was measured every 20 s. The depth-dose and profiles data were collected for various field sizes and depths. For all radiation types and field sizes, the depth-dose data of the diamond chamber showed identical curves to those of the ionization chambers. The profile of the diamond detector was very similar to those of the EDGE and SFD detectors, although the Semiflex and PinPoint chambers showed volume-averaging effects in the penumbrae region. The temperature dependency was within 0.7% in the range of 4-41°C. A dose of 900 cGy and 1200 cGy was needed to stabilize the chamber to the level within 0.5% and 0.2%, respectively. The PTW type 60019 SCDD detector showed suitable characteristics for radiation dosimetry, for relative dose, depth-dose and profile measurements for a wide range of field sizes. However, at least 1000 cGy of pre-irradiation will be needed for accurate measurements.

KW - Characterization

KW - Detector

KW - Dosimetry

KW - Microdiamond

UR - http://www.scopus.com/inward/record.url?scp=84973094050&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84973094050&partnerID=8YFLogxK

U2 - 10.1093/jrr/rrv044

DO - 10.1093/jrr/rrv044

M3 - Article

VL - 56

SP - 912

EP - 918

JO - Journal of Radiation Research

JF - Journal of Radiation Research

SN - 0449-3060

IS - 6

ER -