Development of aorta-to-right atrial fistula following closure of seundum atrial septal defect using the Amplatzer septal occluder

David S. Chun, Mark W. Turrentine, Ahmad Moustapha, Mark H. Hoyer

Research output: Contribution to journalArticle

63 Scopus citations

Abstract

An aorta-to-right atrial fistula developed within 3 months after closure of an atrial septal defect using the Amplatzer septal occluder. Surgical correction was required. The cause of this complication appears to be erosion into the aorta by the right atrial disk.