Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams

I. J. Das, J. C. Harrington, S. F. Akber, A. F. Tomer, J. C. Murray, C. W. Cheng

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

Electron beam dosimetry at low monitor unit (MU) settings is important for dosimetric applications. Dose linearity, beam flatness, and beam energies were studied at low MU settings with various dose rates for different types of linear accelerators. It is observed that for the scattering foil units, the dose/MU is a smooth function of MU for all beam energies. Discrepancies in dose/MU are highest at the lowest MU. Significant variation (5%-245%) in dose linearity is observed among various linear accelerators at low MU settings. Dose rate has no effect on the dose linearity for all energies for the scattering foil units tested. On the contrary, for the scanned beam, there is no predictable pattern as dose/MU is random in nature and varies with time and beam energy. The maximum dosimetric error is observed for the highest energy beam where the beam width is most narrow. Using film, the beam uniformity was noticed to be very poor at low MU and high energy for scanned beams. The beam uniformity and dose linearity are random at low MU due to the random nature of the scan cycle. Under the adverse conditions, the deviation in dosimetric parameters was observed up to 200 MU.

Original languageEnglish (US)
Pages (from-to)821-826
Number of pages6
JournalMedical Physics
Volume21
Issue number6
DOIs
StatePublished - 1994
Externally publishedYes

Fingerprint

Particle Accelerators
Electrons

Keywords

  • dose linearity
  • electron beam
  • end effect
  • monitor unit
  • scanned beam
  • scattering foil

ASJC Scopus subject areas

  • Biophysics

Cite this

Das, I. J., Harrington, J. C., Akber, S. F., Tomer, A. F., Murray, J. C., & Cheng, C. W. (1994). Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams. Medical Physics, 21(6), 821-826. https://doi.org/10.1118/1.597328

Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams. / Das, I. J.; Harrington, J. C.; Akber, S. F.; Tomer, A. F.; Murray, J. C.; Cheng, C. W.

In: Medical Physics, Vol. 21, No. 6, 1994, p. 821-826.

Research output: Contribution to journalArticle

Das, IJ, Harrington, JC, Akber, SF, Tomer, AF, Murray, JC & Cheng, CW 1994, 'Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams', Medical Physics, vol. 21, no. 6, pp. 821-826. https://doi.org/10.1118/1.597328
Das, I. J. ; Harrington, J. C. ; Akber, S. F. ; Tomer, A. F. ; Murray, J. C. ; Cheng, C. W. / Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams. In: Medical Physics. 1994 ; Vol. 21, No. 6. pp. 821-826.
@article{4e509589e4b648578b4722a0a61aa2bf,
title = "Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams",
abstract = "Electron beam dosimetry at low monitor unit (MU) settings is important for dosimetric applications. Dose linearity, beam flatness, and beam energies were studied at low MU settings with various dose rates for different types of linear accelerators. It is observed that for the scattering foil units, the dose/MU is a smooth function of MU for all beam energies. Discrepancies in dose/MU are highest at the lowest MU. Significant variation (5{\%}-245{\%}) in dose linearity is observed among various linear accelerators at low MU settings. Dose rate has no effect on the dose linearity for all energies for the scattering foil units tested. On the contrary, for the scanned beam, there is no predictable pattern as dose/MU is random in nature and varies with time and beam energy. The maximum dosimetric error is observed for the highest energy beam where the beam width is most narrow. Using film, the beam uniformity was noticed to be very poor at low MU and high energy for scanned beams. The beam uniformity and dose linearity are random at low MU due to the random nature of the scan cycle. Under the adverse conditions, the deviation in dosimetric parameters was observed up to 200 MU.",
keywords = "dose linearity, electron beam, end effect, monitor unit, scanned beam, scattering foil",
author = "Das, {I. J.} and Harrington, {J. C.} and Akber, {S. F.} and Tomer, {A. F.} and Murray, {J. C.} and Cheng, {C. W.}",
year = "1994",
doi = "10.1118/1.597328",
language = "English (US)",
volume = "21",
pages = "821--826",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "6",

}

TY - JOUR

T1 - Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beams

AU - Das, I. J.

AU - Harrington, J. C.

AU - Akber, S. F.

AU - Tomer, A. F.

AU - Murray, J. C.

AU - Cheng, C. W.

PY - 1994

Y1 - 1994

N2 - Electron beam dosimetry at low monitor unit (MU) settings is important for dosimetric applications. Dose linearity, beam flatness, and beam energies were studied at low MU settings with various dose rates for different types of linear accelerators. It is observed that for the scattering foil units, the dose/MU is a smooth function of MU for all beam energies. Discrepancies in dose/MU are highest at the lowest MU. Significant variation (5%-245%) in dose linearity is observed among various linear accelerators at low MU settings. Dose rate has no effect on the dose linearity for all energies for the scattering foil units tested. On the contrary, for the scanned beam, there is no predictable pattern as dose/MU is random in nature and varies with time and beam energy. The maximum dosimetric error is observed for the highest energy beam where the beam width is most narrow. Using film, the beam uniformity was noticed to be very poor at low MU and high energy for scanned beams. The beam uniformity and dose linearity are random at low MU due to the random nature of the scan cycle. Under the adverse conditions, the deviation in dosimetric parameters was observed up to 200 MU.

AB - Electron beam dosimetry at low monitor unit (MU) settings is important for dosimetric applications. Dose linearity, beam flatness, and beam energies were studied at low MU settings with various dose rates for different types of linear accelerators. It is observed that for the scattering foil units, the dose/MU is a smooth function of MU for all beam energies. Discrepancies in dose/MU are highest at the lowest MU. Significant variation (5%-245%) in dose linearity is observed among various linear accelerators at low MU settings. Dose rate has no effect on the dose linearity for all energies for the scattering foil units tested. On the contrary, for the scanned beam, there is no predictable pattern as dose/MU is random in nature and varies with time and beam energy. The maximum dosimetric error is observed for the highest energy beam where the beam width is most narrow. Using film, the beam uniformity was noticed to be very poor at low MU and high energy for scanned beams. The beam uniformity and dose linearity are random at low MU due to the random nature of the scan cycle. Under the adverse conditions, the deviation in dosimetric parameters was observed up to 200 MU.

KW - dose linearity

KW - electron beam

KW - end effect

KW - monitor unit

KW - scanned beam

KW - scattering foil

UR - http://www.scopus.com/inward/record.url?scp=0028338920&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0028338920&partnerID=8YFLogxK

U2 - 10.1118/1.597328

DO - 10.1118/1.597328

M3 - Article

C2 - 7935220

AN - SCOPUS:0028338920

VL - 21

SP - 821

EP - 826

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 6

ER -