Abstract
A simple analytical method for the accurate calculation of the relationship between bladder and rectum dose with respect to the source strength is presented. The method can be used for a quick check for the computer-generated results as well as it can be used for simple quality assurance method for dose calculation. This method is accurate to within 3%.
Original language | English (US) |
---|---|
Title of host publication | Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings |
Editors | J.D. Enderle |
Pages | 2771-2773 |
Number of pages | 3 |
Volume | 4 |
State | Published - 2000 |
Externally published | Yes |
Event | 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Chicago, IL, United States Duration: Jul 23 2000 → Jul 28 2000 |
Other
Other | 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society |
---|---|
Country | United States |
City | Chicago, IL |
Period | 7/23/00 → 7/28/00 |
Fingerprint
ASJC Scopus subject areas
- Bioengineering
Cite this
Dosimetric verification of gynecological implants. / Arjomandy, B.; Rajotte, M.; Das, I. J.
Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. ed. / J.D. Enderle. Vol. 4 2000. p. 2771-2773.Research output: Chapter in Book/Report/Conference proceeding › Conference contribution
}
TY - GEN
T1 - Dosimetric verification of gynecological implants
AU - Arjomandy, B.
AU - Rajotte, M.
AU - Das, I. J.
PY - 2000
Y1 - 2000
N2 - A simple analytical method for the accurate calculation of the relationship between bladder and rectum dose with respect to the source strength is presented. The method can be used for a quick check for the computer-generated results as well as it can be used for simple quality assurance method for dose calculation. This method is accurate to within 3%.
AB - A simple analytical method for the accurate calculation of the relationship between bladder and rectum dose with respect to the source strength is presented. The method can be used for a quick check for the computer-generated results as well as it can be used for simple quality assurance method for dose calculation. This method is accurate to within 3%.
UR - http://www.scopus.com/inward/record.url?scp=0034444133&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0034444133&partnerID=8YFLogxK
M3 - Conference contribution
AN - SCOPUS:0034444133
VL - 4
SP - 2771
EP - 2773
BT - Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings
A2 - Enderle, J.D.
ER -