Film dosimetry is a viable alternative to diode array and ion chamber array systems for beam data acquisition of non-physical wedges. Diode array systems exhibit energy dependence in the wedge toe region as shown via comparison of film and Monte Carlo results.

Original languageEnglish (US)
Title of host publicationAnnual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings
EditorsJ.D. Enderle
Pages3144-3147
Number of pages4
Volume4
StatePublished - 2000
Externally publishedYes
Event22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Chicago, IL, United States
Duration: Jul 23 2000Jul 28 2000

Other

Other22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society
CountryUnited States
CityChicago, IL
Period7/23/007/28/00

Fingerprint

Dosimetry
Diodes
Ionization chambers
Data acquisition

ASJC Scopus subject areas

  • Bioengineering

Cite this

Mitra, R. K., Das, I. J., & Verhaegen, F. (2000). Dosimetry and commissioning of virtual wedge using film. In J. D. Enderle (Ed.), Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings (Vol. 4, pp. 3144-3147)

Dosimetry and commissioning of virtual wedge using film. / Mitra, R. K.; Das, I. J.; Verhaegen, F.

Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. ed. / J.D. Enderle. Vol. 4 2000. p. 3144-3147.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Mitra, RK, Das, IJ & Verhaegen, F 2000, Dosimetry and commissioning of virtual wedge using film. in JD Enderle (ed.), Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. vol. 4, pp. 3144-3147, 22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Chicago, IL, United States, 7/23/00.
Mitra RK, Das IJ, Verhaegen F. Dosimetry and commissioning of virtual wedge using film. In Enderle JD, editor, Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. Vol. 4. 2000. p. 3144-3147
Mitra, R. K. ; Das, I. J. ; Verhaegen, F. / Dosimetry and commissioning of virtual wedge using film. Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings. editor / J.D. Enderle. Vol. 4 2000. pp. 3144-3147
@inproceedings{77270850f00d4453a83bfa8a47649e69,
title = "Dosimetry and commissioning of virtual wedge using film",
abstract = "Film dosimetry is a viable alternative to diode array and ion chamber array systems for beam data acquisition of non-physical wedges. Diode array systems exhibit energy dependence in the wedge toe region as shown via comparison of film and Monte Carlo results.",
author = "Mitra, {R. K.} and Das, {I. J.} and F. Verhaegen",
year = "2000",
language = "English (US)",
volume = "4",
pages = "3144--3147",
editor = "J.D. Enderle",
booktitle = "Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings",

}

TY - GEN

T1 - Dosimetry and commissioning of virtual wedge using film

AU - Mitra, R. K.

AU - Das, I. J.

AU - Verhaegen, F.

PY - 2000

Y1 - 2000

N2 - Film dosimetry is a viable alternative to diode array and ion chamber array systems for beam data acquisition of non-physical wedges. Diode array systems exhibit energy dependence in the wedge toe region as shown via comparison of film and Monte Carlo results.

AB - Film dosimetry is a viable alternative to diode array and ion chamber array systems for beam data acquisition of non-physical wedges. Diode array systems exhibit energy dependence in the wedge toe region as shown via comparison of film and Monte Carlo results.

UR - http://www.scopus.com/inward/record.url?scp=0034441561&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034441561&partnerID=8YFLogxK

M3 - Conference contribution

VL - 4

SP - 3144

EP - 3147

BT - Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings

A2 - Enderle, J.D.

ER -