Evaluation of several film/screen combinations for megavoltage radiation therapy treatment field verification

Indra J. Das, Edwin C. McCullough, Francis Weber

Research output: Contribution to journalArticle

Abstract

The introduction of a lower cost alternative to a film commonly used for megavoltage radiation therapy port films led to a comparison of two films (Kodak X-TL and X-OMAT G) in four metal screen cassette configurations. In addition to determination of H-D curves and point spread functions, images of humanoid phantom sections were obtained and evaluated for clinical "usefulness." The X-OMAT G film-screen combinations compared favorably with the X-TL film combinations with respect to the H-D Curves and point spread functions but there was a slight preference for X-TL film. There were no differences noted whether the film was used in the standard "ready pack" or naked in the cassettes.

Original languageEnglish (US)
Pages (from-to)95-97
Number of pages3
JournalMedical Dosimetry
Volume13
Issue number2
StatePublished - 1988
Externally publishedYes

Fingerprint

Motion Pictures
radiation therapy
Radiotherapy
Metals
Costs and Cost Analysis
evaluation
point spread functions
Therapeutics
curves
configurations
metals

Keywords

  • Megavoltage therapy, imaging
  • Megavoltage therapy, port films
  • Screen-film combinations, radiation therapy

ASJC Scopus subject areas

  • Oncology
  • Radiology Nuclear Medicine and imaging
  • Radiation
  • Health Professions(all)
  • Medicine(all)

Cite this

Evaluation of several film/screen combinations for megavoltage radiation therapy treatment field verification. / Das, Indra J.; McCullough, Edwin C.; Weber, Francis.

In: Medical Dosimetry, Vol. 13, No. 2, 1988, p. 95-97.

Research output: Contribution to journalArticle

Das, Indra J. ; McCullough, Edwin C. ; Weber, Francis. / Evaluation of several film/screen combinations for megavoltage radiation therapy treatment field verification. In: Medical Dosimetry. 1988 ; Vol. 13, No. 2. pp. 95-97.
@article{47be2f95a9de423980b35970dc13c139,
title = "Evaluation of several film/screen combinations for megavoltage radiation therapy treatment field verification",
abstract = "The introduction of a lower cost alternative to a film commonly used for megavoltage radiation therapy port films led to a comparison of two films (Kodak X-TL and X-OMAT G) in four metal screen cassette configurations. In addition to determination of H-D curves and point spread functions, images of humanoid phantom sections were obtained and evaluated for clinical {"}usefulness.{"} The X-OMAT G film-screen combinations compared favorably with the X-TL film combinations with respect to the H-D Curves and point spread functions but there was a slight preference for X-TL film. There were no differences noted whether the film was used in the standard {"}ready pack{"} or naked in the cassettes.",
keywords = "Megavoltage therapy, imaging, Megavoltage therapy, port films, Screen-film combinations, radiation therapy",
author = "Das, {Indra J.} and McCullough, {Edwin C.} and Francis Weber",
year = "1988",
language = "English (US)",
volume = "13",
pages = "95--97",
journal = "Medical Dosimetry",
issn = "0958-3947",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Evaluation of several film/screen combinations for megavoltage radiation therapy treatment field verification

AU - Das, Indra J.

AU - McCullough, Edwin C.

AU - Weber, Francis

PY - 1988

Y1 - 1988

N2 - The introduction of a lower cost alternative to a film commonly used for megavoltage radiation therapy port films led to a comparison of two films (Kodak X-TL and X-OMAT G) in four metal screen cassette configurations. In addition to determination of H-D curves and point spread functions, images of humanoid phantom sections were obtained and evaluated for clinical "usefulness." The X-OMAT G film-screen combinations compared favorably with the X-TL film combinations with respect to the H-D Curves and point spread functions but there was a slight preference for X-TL film. There were no differences noted whether the film was used in the standard "ready pack" or naked in the cassettes.

AB - The introduction of a lower cost alternative to a film commonly used for megavoltage radiation therapy port films led to a comparison of two films (Kodak X-TL and X-OMAT G) in four metal screen cassette configurations. In addition to determination of H-D curves and point spread functions, images of humanoid phantom sections were obtained and evaluated for clinical "usefulness." The X-OMAT G film-screen combinations compared favorably with the X-TL film combinations with respect to the H-D Curves and point spread functions but there was a slight preference for X-TL film. There were no differences noted whether the film was used in the standard "ready pack" or naked in the cassettes.

KW - Megavoltage therapy, imaging

KW - Megavoltage therapy, port films

KW - Screen-film combinations, radiation therapy