Evaluation of the quantitative capability of a high-resolution positron emission tomography scanner for small animal imaging

Frederic H. Fahey, H. Donald Gage, Nancy Buchheimer, Holly C. Smith, Beth A. Harkness, Rodney C. Williams, Michael C. Bounds, Jennifer Mercier, Michael E C Robbins, R. Payne, Kathryn A. Morton, Robert H. Mach

Research output: Contribution to journalArticle

15 Citations (Scopus)

Abstract

Objective: The quantitative capability of a positron emission tomography scanner for small animal imaging was evaluated in this study. Methods: The microPET P4 (Concorde Microsystems, Knoxville, TN) scanner's capability for dynamic imaging and corrections for radioactive decay, dead time, and attenuation were evaluated. Rat brain and heart studies with and without attenuation correction were compared. A calibration approach to convert the data to nanocuries per milliliter was implemented. Calibration factors were determined using calibration phantoms of 2 sizes with and without attenuation correction. Quantitation was validated using the MiniPhantom (Data Spectrum, Chapel Hill, NC) with hot features (5:1 ratio) of different sizes (4, 6.4, 8, 13, and 16 mm). Results: The microPET P4 scanner's ability to acquire dynamic studies and to correct for decay, dead time, and attenuation was demonstrated. The microPET P4 scanner provided accurate quantitation to within 6% for features larger than 10 mm. Sixty percent of object contrast was retained for features as small as 4 mm. Conclusions: The microPET P4 scanner can provide accurate quantitation.

Original languageEnglish (US)
Pages (from-to)842-848
Number of pages7
JournalJournal of Computer Assisted Tomography
Volume28
Issue number6
DOIs
StatePublished - Nov 2004
Externally publishedYes

Fingerprint

Positron-Emission Tomography
Calibration
Brain

Keywords

  • Animal imaging
  • Molecular imaging
  • Positron emission tomography
  • Quantitation

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

Evaluation of the quantitative capability of a high-resolution positron emission tomography scanner for small animal imaging. / Fahey, Frederic H.; Gage, H. Donald; Buchheimer, Nancy; Smith, Holly C.; Harkness, Beth A.; Williams, Rodney C.; Bounds, Michael C.; Mercier, Jennifer; Robbins, Michael E C; Payne, R.; Morton, Kathryn A.; Mach, Robert H.

In: Journal of Computer Assisted Tomography, Vol. 28, No. 6, 11.2004, p. 842-848.

Research output: Contribution to journalArticle

Fahey, FH, Gage, HD, Buchheimer, N, Smith, HC, Harkness, BA, Williams, RC, Bounds, MC, Mercier, J, Robbins, MEC, Payne, R, Morton, KA & Mach, RH 2004, 'Evaluation of the quantitative capability of a high-resolution positron emission tomography scanner for small animal imaging', Journal of Computer Assisted Tomography, vol. 28, no. 6, pp. 842-848. https://doi.org/10.1097/00004728-200411000-00020
Fahey, Frederic H. ; Gage, H. Donald ; Buchheimer, Nancy ; Smith, Holly C. ; Harkness, Beth A. ; Williams, Rodney C. ; Bounds, Michael C. ; Mercier, Jennifer ; Robbins, Michael E C ; Payne, R. ; Morton, Kathryn A. ; Mach, Robert H. / Evaluation of the quantitative capability of a high-resolution positron emission tomography scanner for small animal imaging. In: Journal of Computer Assisted Tomography. 2004 ; Vol. 28, No. 6. pp. 842-848.
@article{480ab9872c5f47acae5b984794c552dd,
title = "Evaluation of the quantitative capability of a high-resolution positron emission tomography scanner for small animal imaging",
abstract = "Objective: The quantitative capability of a positron emission tomography scanner for small animal imaging was evaluated in this study. Methods: The microPET P4 (Concorde Microsystems, Knoxville, TN) scanner's capability for dynamic imaging and corrections for radioactive decay, dead time, and attenuation were evaluated. Rat brain and heart studies with and without attenuation correction were compared. A calibration approach to convert the data to nanocuries per milliliter was implemented. Calibration factors were determined using calibration phantoms of 2 sizes with and without attenuation correction. Quantitation was validated using the MiniPhantom (Data Spectrum, Chapel Hill, NC) with hot features (5:1 ratio) of different sizes (4, 6.4, 8, 13, and 16 mm). Results: The microPET P4 scanner's ability to acquire dynamic studies and to correct for decay, dead time, and attenuation was demonstrated. The microPET P4 scanner provided accurate quantitation to within 6{\%} for features larger than 10 mm. Sixty percent of object contrast was retained for features as small as 4 mm. Conclusions: The microPET P4 scanner can provide accurate quantitation.",
keywords = "Animal imaging, Molecular imaging, Positron emission tomography, Quantitation",
author = "Fahey, {Frederic H.} and Gage, {H. Donald} and Nancy Buchheimer and Smith, {Holly C.} and Harkness, {Beth A.} and Williams, {Rodney C.} and Bounds, {Michael C.} and Jennifer Mercier and Robbins, {Michael E C} and R. Payne and Morton, {Kathryn A.} and Mach, {Robert H.}",
year = "2004",
month = "11",
doi = "10.1097/00004728-200411000-00020",
language = "English (US)",
volume = "28",
pages = "842--848",
journal = "Journal of Computer Assisted Tomography",
issn = "0363-8715",
publisher = "Lippincott Williams and Wilkins",
number = "6",

}

TY - JOUR

T1 - Evaluation of the quantitative capability of a high-resolution positron emission tomography scanner for small animal imaging

AU - Fahey, Frederic H.

AU - Gage, H. Donald

AU - Buchheimer, Nancy

AU - Smith, Holly C.

AU - Harkness, Beth A.

AU - Williams, Rodney C.

AU - Bounds, Michael C.

AU - Mercier, Jennifer

AU - Robbins, Michael E C

AU - Payne, R.

AU - Morton, Kathryn A.

AU - Mach, Robert H.

PY - 2004/11

Y1 - 2004/11

N2 - Objective: The quantitative capability of a positron emission tomography scanner for small animal imaging was evaluated in this study. Methods: The microPET P4 (Concorde Microsystems, Knoxville, TN) scanner's capability for dynamic imaging and corrections for radioactive decay, dead time, and attenuation were evaluated. Rat brain and heart studies with and without attenuation correction were compared. A calibration approach to convert the data to nanocuries per milliliter was implemented. Calibration factors were determined using calibration phantoms of 2 sizes with and without attenuation correction. Quantitation was validated using the MiniPhantom (Data Spectrum, Chapel Hill, NC) with hot features (5:1 ratio) of different sizes (4, 6.4, 8, 13, and 16 mm). Results: The microPET P4 scanner's ability to acquire dynamic studies and to correct for decay, dead time, and attenuation was demonstrated. The microPET P4 scanner provided accurate quantitation to within 6% for features larger than 10 mm. Sixty percent of object contrast was retained for features as small as 4 mm. Conclusions: The microPET P4 scanner can provide accurate quantitation.

AB - Objective: The quantitative capability of a positron emission tomography scanner for small animal imaging was evaluated in this study. Methods: The microPET P4 (Concorde Microsystems, Knoxville, TN) scanner's capability for dynamic imaging and corrections for radioactive decay, dead time, and attenuation were evaluated. Rat brain and heart studies with and without attenuation correction were compared. A calibration approach to convert the data to nanocuries per milliliter was implemented. Calibration factors were determined using calibration phantoms of 2 sizes with and without attenuation correction. Quantitation was validated using the MiniPhantom (Data Spectrum, Chapel Hill, NC) with hot features (5:1 ratio) of different sizes (4, 6.4, 8, 13, and 16 mm). Results: The microPET P4 scanner's ability to acquire dynamic studies and to correct for decay, dead time, and attenuation was demonstrated. The microPET P4 scanner provided accurate quantitation to within 6% for features larger than 10 mm. Sixty percent of object contrast was retained for features as small as 4 mm. Conclusions: The microPET P4 scanner can provide accurate quantitation.

KW - Animal imaging

KW - Molecular imaging

KW - Positron emission tomography

KW - Quantitation

UR - http://www.scopus.com/inward/record.url?scp=8444220899&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=8444220899&partnerID=8YFLogxK

U2 - 10.1097/00004728-200411000-00020

DO - 10.1097/00004728-200411000-00020

M3 - Article

C2 - 15538162

AN - SCOPUS:8444220899

VL - 28

SP - 842

EP - 848

JO - Journal of Computer Assisted Tomography

JF - Journal of Computer Assisted Tomography

SN - 0363-8715

IS - 6

ER -