In vitro and in vivo measurement of red cell velocity with epi- and transillumination

Scot L. Harper, H. Glenn Bohlen

Research output: Contribution to journalArticle

11 Scopus citations

Abstract

Measurement of red cell velocity with the dual-slit cross-correlation method in glass capillary tubes during transillumination indicates that the measured velocity must be divided by a correction factor of approximately 1.6 to equal the average velocity calculated from a known flow and inner diameter. Whether the same correction factor exists when red cell velocity is measured during epiillumination is questionable. Red cell velocity was measured with the dual-slit correlation method nearly simultaneously using epi- (EL) and transillumination (TL) while glass tubes (40-100 μm, i.d.) were pump perfused with whole human blood (hemotocrit 39-42%). With TL, the measured velocity is 1.58 ± 0.07 (SEM) times the calculated average velocity, whereas a factor of 2.04 ± 0.04 (SEM) was obtained with epiillumination. When intestinal arterioles with approximately the same inner diameters and flow velocities as the glass tubes were used, the ratio of velocities measured with TL to EL was 1.21 ± 0.02 (SEM) as compared to 1.31 ± 0.09 (SEM) for glass tubes using TL and EL of the tube at the same pump flow. This similarity of TL to EL velocity ratios for glass tubes and microvessels may be fortuitous or indicate that comparable flow properties and measurement conditions exist for in vitro and in vivo situations. The major finding of the study is, however, that different velocity correction factors exist for EL and TL measurements when the dual-slit correlation method is used to estimate red cell velocities in tubes of an internal diameter of 40-100 μm at normal hematocrits.

Original languageEnglish (US)
Pages (from-to)186-193
Number of pages8
JournalMicrovascular Research
Volume25
Issue number2
DOIs
StatePublished - Mar 1983

ASJC Scopus subject areas

  • Biochemistry
  • Cardiology and Cardiovascular Medicine
  • Cell Biology

Fingerprint Dive into the research topics of 'In vitro and in vivo measurement of red cell velocity with epi- and transillumination'. Together they form a unique fingerprint.

  • Cite this