Increased analytical precision in the hollow cathode discharge emission source by improved discharge current control

Jih Lie Tseng, J. C. Williams, Robert B. Bartlow, Steven T. Griffin, James Williams

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

A significant improvement in the precision of the hollow cathode as an emission source is reported. Precision of 1% or less has been observed several times over periods of several hours. An average long-term stability of 4.3% for Li and 6.0% for Na in the emission signal from microsamples (

Original languageEnglish (US)
Pages (from-to)1933-1942
Number of pages10
JournalAnalytical Chemistry
Volume63
Issue number18
StatePublished - Sep 15 1991
Externally publishedYes

Fingerprint

Electric current control
Cathodes

ASJC Scopus subject areas

  • Analytical Chemistry

Cite this

Increased analytical precision in the hollow cathode discharge emission source by improved discharge current control. / Tseng, Jih Lie; Williams, J. C.; Bartlow, Robert B.; Griffin, Steven T.; Williams, James.

In: Analytical Chemistry, Vol. 63, No. 18, 15.09.1991, p. 1933-1942.

Research output: Contribution to journalArticle

Tseng, Jih Lie ; Williams, J. C. ; Bartlow, Robert B. ; Griffin, Steven T. ; Williams, James. / Increased analytical precision in the hollow cathode discharge emission source by improved discharge current control. In: Analytical Chemistry. 1991 ; Vol. 63, No. 18. pp. 1933-1942.
@article{61eac00682464862b6df99f5f2d98030,
title = "Increased analytical precision in the hollow cathode discharge emission source by improved discharge current control",
abstract = "A significant improvement in the precision of the hollow cathode as an emission source is reported. Precision of 1{\%} or less has been observed several times over periods of several hours. An average long-term stability of 4.3{\%} for Li and 6.0{\%} for Na in the emission signal from microsamples (",
author = "Tseng, {Jih Lie} and Williams, {J. C.} and Bartlow, {Robert B.} and Griffin, {Steven T.} and James Williams",
year = "1991",
month = "9",
day = "15",
language = "English (US)",
volume = "63",
pages = "1933--1942",
journal = "Analytical Chemistry",
issn = "0003-2700",
publisher = "American Chemical Society",
number = "18",

}

TY - JOUR

T1 - Increased analytical precision in the hollow cathode discharge emission source by improved discharge current control

AU - Tseng, Jih Lie

AU - Williams, J. C.

AU - Bartlow, Robert B.

AU - Griffin, Steven T.

AU - Williams, James

PY - 1991/9/15

Y1 - 1991/9/15

N2 - A significant improvement in the precision of the hollow cathode as an emission source is reported. Precision of 1% or less has been observed several times over periods of several hours. An average long-term stability of 4.3% for Li and 6.0% for Na in the emission signal from microsamples (

AB - A significant improvement in the precision of the hollow cathode as an emission source is reported. Precision of 1% or less has been observed several times over periods of several hours. An average long-term stability of 4.3% for Li and 6.0% for Na in the emission signal from microsamples (

UR - http://www.scopus.com/inward/record.url?scp=0026425533&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0026425533&partnerID=8YFLogxK

M3 - Article

VL - 63

SP - 1933

EP - 1942

JO - Analytical Chemistry

JF - Analytical Chemistry

SN - 0003-2700

IS - 18