Inner ear endoscopy during cochlear implantation of cochlear dysplasia

M. H. Fritsch

Research output: Contribution to journalArticle

Abstract

implantation by visualizing the dysplastic inner ear in order to visually guide the process of cochlear implant electrode insertion. A retrospective case review was conducted in a tertiary referral center. Three cochlear dysplasia patients representing five ears were endoscopically operated for cochlear implantation. Endoscopes (telescopic and fiberoptic) were used during cochlear implant surgery to visualize dysplastic inner ear structures and to guide the implant electrodes through the abnormal inner ear anatomy in real-time to their final best resting place within the cochlea. Using two endoscope types and the microscope for viewing the inner ear and inserting the implant electrodes, the ability to visualize anatomy and actively guide the electrodes was assessed. The five ears operated upon showed that endoscopic viewing of the dysplastic inner ear was superior to microscope use. All five ears showed different intra-cochlear anatomy. Some of the findings were: Each ear showed a foreshortened Scala tympani; fluid pulsations of various strengths representing cerebrospinal fluid were seen in all ears; all scalae were in communication with each other without developed partitions between them. No inner ear anatomy relating to functional hearing was seen. New surgically relevant anatomy was seen that altered surgical cochlear implantation technique during the case. Endoscopically, intra-cochlear anatomic relationships can be seen in real-time in three dimensions. Each dysplastic ear of this series had different anatomy representing a different end-point of development. No residual hearing anatomy was seen in any of the dysplastic ears. For patients with both profound hearing loss and cochlear dysplasia, the absence of anatomy seen in these five ears predicts that a hearing aid trial would be unsuccessful. In these audiometrically verified profound hearing loss cochlear implant candidates with dysplastic cochleas, early implantation without a hearing aid trial is encouraged. Using an endoscope, the cochlear implant electrode can be directly viewed and positioned during surgery using an endoscope. Due to the foreshortened length of the cochlear spiral, a compressed array electrode may be the best option for some cochlear deformity ears. New electrodes specifically designed for the common cavity ear are needed.

Original languageEnglish
Pages (from-to)205-212
Number of pages8
JournalOtorinolaringologia
Volume61
Issue number4
StatePublished - Dec 2011

Fingerprint

Cochlear Implantation
Cochlea
Inner Ear
Endoscopy
Ear
Anatomy
Electrodes
Cochlear Implants
Endoscopes
Hearing Aids
Hearing
Scala Tympani
Aptitude
Sensorineural Hearing Loss
Hearing Loss
Tertiary Care Centers
Cerebrospinal Fluid
Communication

Keywords

  • Cochlea
  • Ear
  • Endoscopy
  • Inner
  • Operative
  • Surgical procedures

ASJC Scopus subject areas

  • Otorhinolaryngology

Cite this

Inner ear endoscopy during cochlear implantation of cochlear dysplasia. / Fritsch, M. H.

In: Otorinolaringologia, Vol. 61, No. 4, 12.2011, p. 205-212.

Research output: Contribution to journalArticle

Fritsch, M. H. / Inner ear endoscopy during cochlear implantation of cochlear dysplasia. In: Otorinolaringologia. 2011 ; Vol. 61, No. 4. pp. 205-212.
@article{9d8dc1a4e7774f72bc20f0ac3543a637,
title = "Inner ear endoscopy during cochlear implantation of cochlear dysplasia",
abstract = "implantation by visualizing the dysplastic inner ear in order to visually guide the process of cochlear implant electrode insertion. A retrospective case review was conducted in a tertiary referral center. Three cochlear dysplasia patients representing five ears were endoscopically operated for cochlear implantation. Endoscopes (telescopic and fiberoptic) were used during cochlear implant surgery to visualize dysplastic inner ear structures and to guide the implant electrodes through the abnormal inner ear anatomy in real-time to their final best resting place within the cochlea. Using two endoscope types and the microscope for viewing the inner ear and inserting the implant electrodes, the ability to visualize anatomy and actively guide the electrodes was assessed. The five ears operated upon showed that endoscopic viewing of the dysplastic inner ear was superior to microscope use. All five ears showed different intra-cochlear anatomy. Some of the findings were: Each ear showed a foreshortened Scala tympani; fluid pulsations of various strengths representing cerebrospinal fluid were seen in all ears; all scalae were in communication with each other without developed partitions between them. No inner ear anatomy relating to functional hearing was seen. New surgically relevant anatomy was seen that altered surgical cochlear implantation technique during the case. Endoscopically, intra-cochlear anatomic relationships can be seen in real-time in three dimensions. Each dysplastic ear of this series had different anatomy representing a different end-point of development. No residual hearing anatomy was seen in any of the dysplastic ears. For patients with both profound hearing loss and cochlear dysplasia, the absence of anatomy seen in these five ears predicts that a hearing aid trial would be unsuccessful. In these audiometrically verified profound hearing loss cochlear implant candidates with dysplastic cochleas, early implantation without a hearing aid trial is encouraged. Using an endoscope, the cochlear implant electrode can be directly viewed and positioned during surgery using an endoscope. Due to the foreshortened length of the cochlear spiral, a compressed array electrode may be the best option for some cochlear deformity ears. New electrodes specifically designed for the common cavity ear are needed.",
keywords = "Cochlea, Ear, Endoscopy, Inner, Operative, Surgical procedures",
author = "Fritsch, {M. H.}",
year = "2011",
month = "12",
language = "English",
volume = "61",
pages = "205--212",
journal = "Otorinolaringologia",
issn = "0392-6621",
publisher = "Asociatia Medicala Romana",
number = "4",

}

TY - JOUR

T1 - Inner ear endoscopy during cochlear implantation of cochlear dysplasia

AU - Fritsch, M. H.

PY - 2011/12

Y1 - 2011/12

N2 - implantation by visualizing the dysplastic inner ear in order to visually guide the process of cochlear implant electrode insertion. A retrospective case review was conducted in a tertiary referral center. Three cochlear dysplasia patients representing five ears were endoscopically operated for cochlear implantation. Endoscopes (telescopic and fiberoptic) were used during cochlear implant surgery to visualize dysplastic inner ear structures and to guide the implant electrodes through the abnormal inner ear anatomy in real-time to their final best resting place within the cochlea. Using two endoscope types and the microscope for viewing the inner ear and inserting the implant electrodes, the ability to visualize anatomy and actively guide the electrodes was assessed. The five ears operated upon showed that endoscopic viewing of the dysplastic inner ear was superior to microscope use. All five ears showed different intra-cochlear anatomy. Some of the findings were: Each ear showed a foreshortened Scala tympani; fluid pulsations of various strengths representing cerebrospinal fluid were seen in all ears; all scalae were in communication with each other without developed partitions between them. No inner ear anatomy relating to functional hearing was seen. New surgically relevant anatomy was seen that altered surgical cochlear implantation technique during the case. Endoscopically, intra-cochlear anatomic relationships can be seen in real-time in three dimensions. Each dysplastic ear of this series had different anatomy representing a different end-point of development. No residual hearing anatomy was seen in any of the dysplastic ears. For patients with both profound hearing loss and cochlear dysplasia, the absence of anatomy seen in these five ears predicts that a hearing aid trial would be unsuccessful. In these audiometrically verified profound hearing loss cochlear implant candidates with dysplastic cochleas, early implantation without a hearing aid trial is encouraged. Using an endoscope, the cochlear implant electrode can be directly viewed and positioned during surgery using an endoscope. Due to the foreshortened length of the cochlear spiral, a compressed array electrode may be the best option for some cochlear deformity ears. New electrodes specifically designed for the common cavity ear are needed.

AB - implantation by visualizing the dysplastic inner ear in order to visually guide the process of cochlear implant electrode insertion. A retrospective case review was conducted in a tertiary referral center. Three cochlear dysplasia patients representing five ears were endoscopically operated for cochlear implantation. Endoscopes (telescopic and fiberoptic) were used during cochlear implant surgery to visualize dysplastic inner ear structures and to guide the implant electrodes through the abnormal inner ear anatomy in real-time to their final best resting place within the cochlea. Using two endoscope types and the microscope for viewing the inner ear and inserting the implant electrodes, the ability to visualize anatomy and actively guide the electrodes was assessed. The five ears operated upon showed that endoscopic viewing of the dysplastic inner ear was superior to microscope use. All five ears showed different intra-cochlear anatomy. Some of the findings were: Each ear showed a foreshortened Scala tympani; fluid pulsations of various strengths representing cerebrospinal fluid were seen in all ears; all scalae were in communication with each other without developed partitions between them. No inner ear anatomy relating to functional hearing was seen. New surgically relevant anatomy was seen that altered surgical cochlear implantation technique during the case. Endoscopically, intra-cochlear anatomic relationships can be seen in real-time in three dimensions. Each dysplastic ear of this series had different anatomy representing a different end-point of development. No residual hearing anatomy was seen in any of the dysplastic ears. For patients with both profound hearing loss and cochlear dysplasia, the absence of anatomy seen in these five ears predicts that a hearing aid trial would be unsuccessful. In these audiometrically verified profound hearing loss cochlear implant candidates with dysplastic cochleas, early implantation without a hearing aid trial is encouraged. Using an endoscope, the cochlear implant electrode can be directly viewed and positioned during surgery using an endoscope. Due to the foreshortened length of the cochlear spiral, a compressed array electrode may be the best option for some cochlear deformity ears. New electrodes specifically designed for the common cavity ear are needed.

KW - Cochlea

KW - Ear

KW - Endoscopy

KW - Inner

KW - Operative

KW - Surgical procedures

UR - http://www.scopus.com/inward/record.url?scp=84856330590&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84856330590&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:84856330590

VL - 61

SP - 205

EP - 212

JO - Otorinolaringologia

JF - Otorinolaringologia

SN - 0392-6621

IS - 4

ER -