Laparoscopic partial kidney ablation with high intensity focused ultrasound

Ryan F. Paterson, Eric Barret, Tibério M. Siqueira, Thomas A. Gardner, Jahangir Tavakkoli, Victor V. Rao, Narendra T. Sanghvi, Liang Cheng, Arieh L. Shalhav

Research output: Contribution to journalArticle

44 Citations (Scopus)

Abstract

Purpose: High intensity focused ultrasound has been performed for transrectal and extracorporeal thermal ablation of tissues. We developed and tested a laparoscopic probe that allows real-time ultrasound imaging during partial renal ablation using high intensity focused ultrasound. Methods: A Sonablate 200 (Focus Surgery, Indianapolis, Indiana) high intensity focused ultrasound system with a modified 18 mm. laparoscopic probe was used in all experiments. In 13 Yucatan mini-pigs a 5Fr ureteral catheter was inserted into the renal pelvis and 10 cc air were instilled into the collecting system. The kidney was laparoscopically dissected, the high intensity focused ultrasound probe was inserted through a 33 mm. laparoscopic port and the targeted renal pole was treated. Results: Renal lesions were created in 12 of 13 treated kidneys under real-time ultrasound visualization. Median operative time was 180 minutes, average high intensity focused ultrasound activation time was 18.3 minutes and lesion size was 23 x 17 x 11 mm. At 4 and 14 days 4 (acute group) and 6 (subacute group) animals were available for renal functional and anatomical evaluation, respectively. No difference in renal function was seen in treated and untreated kidneys. Pathological examination at 14 days revealed homogenous and complete tissue necrosis throughout the whole volume of the lesion with sharp demarcation from adjacent normal tissue. Conclusions: We were able to refine a probe for laparoscopic high intensity focused ultrasound delivery capable of simultaneous ultrasound imaging. Partial renal ablation using this probe is feasible and safe, and resulted in homogenous, complete and reproducible lesions.

Original languageEnglish (US)
Pages (from-to)347-351
Number of pages5
JournalJournal of Urology
Volume169
Issue number1
DOIs
StatePublished - Jan 1 2003

Fingerprint

High-Intensity Focused Ultrasound Ablation
Kidney
Ultrasonography
Urinary Catheters
Kidney Pelvis
Operative Time
Necrosis
Swine
Hot Temperature

Keywords

  • Equipment and supplies
  • Kidney
  • Laparoscopy
  • Swine, miniature
  • Ultrasonic therapy

ASJC Scopus subject areas

  • Urology

Cite this

Paterson, R. F., Barret, E., Siqueira, T. M., Gardner, T. A., Tavakkoli, J., Rao, V. V., ... Shalhav, A. L. (2003). Laparoscopic partial kidney ablation with high intensity focused ultrasound. Journal of Urology, 169(1), 347-351. https://doi.org/10.1016/S0022-5347(05)64124-9

Laparoscopic partial kidney ablation with high intensity focused ultrasound. / Paterson, Ryan F.; Barret, Eric; Siqueira, Tibério M.; Gardner, Thomas A.; Tavakkoli, Jahangir; Rao, Victor V.; Sanghvi, Narendra T.; Cheng, Liang; Shalhav, Arieh L.

In: Journal of Urology, Vol. 169, No. 1, 01.01.2003, p. 347-351.

Research output: Contribution to journalArticle

Paterson, RF, Barret, E, Siqueira, TM, Gardner, TA, Tavakkoli, J, Rao, VV, Sanghvi, NT, Cheng, L & Shalhav, AL 2003, 'Laparoscopic partial kidney ablation with high intensity focused ultrasound', Journal of Urology, vol. 169, no. 1, pp. 347-351. https://doi.org/10.1016/S0022-5347(05)64124-9
Paterson, Ryan F. ; Barret, Eric ; Siqueira, Tibério M. ; Gardner, Thomas A. ; Tavakkoli, Jahangir ; Rao, Victor V. ; Sanghvi, Narendra T. ; Cheng, Liang ; Shalhav, Arieh L. / Laparoscopic partial kidney ablation with high intensity focused ultrasound. In: Journal of Urology. 2003 ; Vol. 169, No. 1. pp. 347-351.
@article{17684c6521214b4d80cf1e4083894e2c,
title = "Laparoscopic partial kidney ablation with high intensity focused ultrasound",
abstract = "Purpose: High intensity focused ultrasound has been performed for transrectal and extracorporeal thermal ablation of tissues. We developed and tested a laparoscopic probe that allows real-time ultrasound imaging during partial renal ablation using high intensity focused ultrasound. Methods: A Sonablate 200 (Focus Surgery, Indianapolis, Indiana) high intensity focused ultrasound system with a modified 18 mm. laparoscopic probe was used in all experiments. In 13 Yucatan mini-pigs a 5Fr ureteral catheter was inserted into the renal pelvis and 10 cc air were instilled into the collecting system. The kidney was laparoscopically dissected, the high intensity focused ultrasound probe was inserted through a 33 mm. laparoscopic port and the targeted renal pole was treated. Results: Renal lesions were created in 12 of 13 treated kidneys under real-time ultrasound visualization. Median operative time was 180 minutes, average high intensity focused ultrasound activation time was 18.3 minutes and lesion size was 23 x 17 x 11 mm. At 4 and 14 days 4 (acute group) and 6 (subacute group) animals were available for renal functional and anatomical evaluation, respectively. No difference in renal function was seen in treated and untreated kidneys. Pathological examination at 14 days revealed homogenous and complete tissue necrosis throughout the whole volume of the lesion with sharp demarcation from adjacent normal tissue. Conclusions: We were able to refine a probe for laparoscopic high intensity focused ultrasound delivery capable of simultaneous ultrasound imaging. Partial renal ablation using this probe is feasible and safe, and resulted in homogenous, complete and reproducible lesions.",
keywords = "Equipment and supplies, Kidney, Laparoscopy, Swine, miniature, Ultrasonic therapy",
author = "Paterson, {Ryan F.} and Eric Barret and Siqueira, {Tib{\'e}rio M.} and Gardner, {Thomas A.} and Jahangir Tavakkoli and Rao, {Victor V.} and Sanghvi, {Narendra T.} and Liang Cheng and Shalhav, {Arieh L.}",
year = "2003",
month = "1",
day = "1",
doi = "10.1016/S0022-5347(05)64124-9",
language = "English (US)",
volume = "169",
pages = "347--351",
journal = "Journal of Urology",
issn = "0022-5347",
publisher = "Elsevier Inc.",
number = "1",

}

TY - JOUR

T1 - Laparoscopic partial kidney ablation with high intensity focused ultrasound

AU - Paterson, Ryan F.

AU - Barret, Eric

AU - Siqueira, Tibério M.

AU - Gardner, Thomas A.

AU - Tavakkoli, Jahangir

AU - Rao, Victor V.

AU - Sanghvi, Narendra T.

AU - Cheng, Liang

AU - Shalhav, Arieh L.

PY - 2003/1/1

Y1 - 2003/1/1

N2 - Purpose: High intensity focused ultrasound has been performed for transrectal and extracorporeal thermal ablation of tissues. We developed and tested a laparoscopic probe that allows real-time ultrasound imaging during partial renal ablation using high intensity focused ultrasound. Methods: A Sonablate 200 (Focus Surgery, Indianapolis, Indiana) high intensity focused ultrasound system with a modified 18 mm. laparoscopic probe was used in all experiments. In 13 Yucatan mini-pigs a 5Fr ureteral catheter was inserted into the renal pelvis and 10 cc air were instilled into the collecting system. The kidney was laparoscopically dissected, the high intensity focused ultrasound probe was inserted through a 33 mm. laparoscopic port and the targeted renal pole was treated. Results: Renal lesions were created in 12 of 13 treated kidneys under real-time ultrasound visualization. Median operative time was 180 minutes, average high intensity focused ultrasound activation time was 18.3 minutes and lesion size was 23 x 17 x 11 mm. At 4 and 14 days 4 (acute group) and 6 (subacute group) animals were available for renal functional and anatomical evaluation, respectively. No difference in renal function was seen in treated and untreated kidneys. Pathological examination at 14 days revealed homogenous and complete tissue necrosis throughout the whole volume of the lesion with sharp demarcation from adjacent normal tissue. Conclusions: We were able to refine a probe for laparoscopic high intensity focused ultrasound delivery capable of simultaneous ultrasound imaging. Partial renal ablation using this probe is feasible and safe, and resulted in homogenous, complete and reproducible lesions.

AB - Purpose: High intensity focused ultrasound has been performed for transrectal and extracorporeal thermal ablation of tissues. We developed and tested a laparoscopic probe that allows real-time ultrasound imaging during partial renal ablation using high intensity focused ultrasound. Methods: A Sonablate 200 (Focus Surgery, Indianapolis, Indiana) high intensity focused ultrasound system with a modified 18 mm. laparoscopic probe was used in all experiments. In 13 Yucatan mini-pigs a 5Fr ureteral catheter was inserted into the renal pelvis and 10 cc air were instilled into the collecting system. The kidney was laparoscopically dissected, the high intensity focused ultrasound probe was inserted through a 33 mm. laparoscopic port and the targeted renal pole was treated. Results: Renal lesions were created in 12 of 13 treated kidneys under real-time ultrasound visualization. Median operative time was 180 minutes, average high intensity focused ultrasound activation time was 18.3 minutes and lesion size was 23 x 17 x 11 mm. At 4 and 14 days 4 (acute group) and 6 (subacute group) animals were available for renal functional and anatomical evaluation, respectively. No difference in renal function was seen in treated and untreated kidneys. Pathological examination at 14 days revealed homogenous and complete tissue necrosis throughout the whole volume of the lesion with sharp demarcation from adjacent normal tissue. Conclusions: We were able to refine a probe for laparoscopic high intensity focused ultrasound delivery capable of simultaneous ultrasound imaging. Partial renal ablation using this probe is feasible and safe, and resulted in homogenous, complete and reproducible lesions.

KW - Equipment and supplies

KW - Kidney

KW - Laparoscopy

KW - Swine, miniature

KW - Ultrasonic therapy

UR - http://www.scopus.com/inward/record.url?scp=0037213191&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0037213191&partnerID=8YFLogxK

U2 - 10.1016/S0022-5347(05)64124-9

DO - 10.1016/S0022-5347(05)64124-9

M3 - Article

C2 - 12478187

AN - SCOPUS:0037213191

VL - 169

SP - 347

EP - 351

JO - Journal of Urology

JF - Journal of Urology

SN - 0022-5347

IS - 1

ER -