Microdamage adjacent to endosseous implants

S. S. Huja, T. R. Katona, D. B. Burr, L. P. Garetto, W. E. Roberts

Research output: Contribution to journalArticle

66 Citations (Scopus)

Abstract

Intense remodeling occurs in lamellar bone adjacent to osseointegrated endosseous implants. The purpose of this study was to compare microdamage accumulation subsequent to ex vivo fatigue loading of bone that surrounds an endosseous implant, (a) immediately after placement (nonadapted bone) and (b) following a 12 week healing period after placement (adapted bone). We hypothesize that there is less microdamage in the more compliant adapted bone than in the older nonadapted bone. Nonthreaded titanium plasma sprayed (TPS)-coated endosseous implants were placed into dog midfemoral diaphyses and allowed to heal for 12 weeks. Block sections of bone, each containing one implant, were cut anteroposteriorly, resulting in an implant containing lateral cortex, and a medial cortex that was used for testing the nonadapted specimens. Control specimens (n = 14 each for adapted and nonadapted) were loaded at 0 N. Experimental specimens (n = 13, adapted; n = 14, nonadapted) were loaded at 100 N in cantilever bending for 150,000 cycles at 2 Hz, at 37°C on a Bionix 858 testing machine. Specimens were bulk stained with basic fuchsin and 120-140 μm sections were obtained. Crack numerical density (Cr.Dn = Cr.N/B.Ar, #/mm2), crack surface density (Cr.S.Dn = Tt.Cr.Le/B.Ar, mm/mm2), and percent damage area (Dm.Ar = Cr.Ar x 100/B.Ar, mm2/mm2) were measured at x250. Statistically significant differences (p < 0.0001) were seen for Cr.Dn, Cr.S.Dn, and Dm.Ar on the compressed cortices suggesting that adapted bone near the implant accumulated significantly less microdamage than nonadapted bone. Also, the adapted nonloaded control specimens had approximately 20-fold less damage than the respective nonadapted specimens. This study suggests that the compliant adapted bone adjacent to endosseous implants is relatively resistant to fatigue loads. The high success rates of endosseous implants may be due to the presence of a rapidly remodeling region that maintains tissue compliance and limits microdamage initiation. Copyright (C) 1999 Elsevier Science Inc.

Original languageEnglish (US)
Pages (from-to)217-222
Number of pages6
JournalBone
Volume25
Issue number2
DOIs
StatePublished - Aug 1999

Fingerprint

Bone and Bones
Fatigue
Diaphyses
Titanium
Compliance
Dogs

Keywords

  • Bone
  • Endosseous implants
  • Fatigue
  • Histomorphometry
  • Microdamage
  • Remodeling

ASJC Scopus subject areas

  • Physiology
  • Hematology

Cite this

Huja, S. S., Katona, T. R., Burr, D. B., Garetto, L. P., & Roberts, W. E. (1999). Microdamage adjacent to endosseous implants. Bone, 25(2), 217-222. https://doi.org/10.1016/S8756-3282(99)00151-9

Microdamage adjacent to endosseous implants. / Huja, S. S.; Katona, T. R.; Burr, D. B.; Garetto, L. P.; Roberts, W. E.

In: Bone, Vol. 25, No. 2, 08.1999, p. 217-222.

Research output: Contribution to journalArticle

Huja, SS, Katona, TR, Burr, DB, Garetto, LP & Roberts, WE 1999, 'Microdamage adjacent to endosseous implants', Bone, vol. 25, no. 2, pp. 217-222. https://doi.org/10.1016/S8756-3282(99)00151-9
Huja SS, Katona TR, Burr DB, Garetto LP, Roberts WE. Microdamage adjacent to endosseous implants. Bone. 1999 Aug;25(2):217-222. https://doi.org/10.1016/S8756-3282(99)00151-9
Huja, S. S. ; Katona, T. R. ; Burr, D. B. ; Garetto, L. P. ; Roberts, W. E. / Microdamage adjacent to endosseous implants. In: Bone. 1999 ; Vol. 25, No. 2. pp. 217-222.
@article{3ab62d488e8846319945bf3fd37df06b,
title = "Microdamage adjacent to endosseous implants",
abstract = "Intense remodeling occurs in lamellar bone adjacent to osseointegrated endosseous implants. The purpose of this study was to compare microdamage accumulation subsequent to ex vivo fatigue loading of bone that surrounds an endosseous implant, (a) immediately after placement (nonadapted bone) and (b) following a 12 week healing period after placement (adapted bone). We hypothesize that there is less microdamage in the more compliant adapted bone than in the older nonadapted bone. Nonthreaded titanium plasma sprayed (TPS)-coated endosseous implants were placed into dog midfemoral diaphyses and allowed to heal for 12 weeks. Block sections of bone, each containing one implant, were cut anteroposteriorly, resulting in an implant containing lateral cortex, and a medial cortex that was used for testing the nonadapted specimens. Control specimens (n = 14 each for adapted and nonadapted) were loaded at 0 N. Experimental specimens (n = 13, adapted; n = 14, nonadapted) were loaded at 100 N in cantilever bending for 150,000 cycles at 2 Hz, at 37°C on a Bionix 858 testing machine. Specimens were bulk stained with basic fuchsin and 120-140 μm sections were obtained. Crack numerical density (Cr.Dn = Cr.N/B.Ar, #/mm2), crack surface density (Cr.S.Dn = Tt.Cr.Le/B.Ar, mm/mm2), and percent damage area (Dm.Ar = Cr.Ar x 100/B.Ar, mm2/mm2) were measured at x250. Statistically significant differences (p < 0.0001) were seen for Cr.Dn, Cr.S.Dn, and Dm.Ar on the compressed cortices suggesting that adapted bone near the implant accumulated significantly less microdamage than nonadapted bone. Also, the adapted nonloaded control specimens had approximately 20-fold less damage than the respective nonadapted specimens. This study suggests that the compliant adapted bone adjacent to endosseous implants is relatively resistant to fatigue loads. The high success rates of endosseous implants may be due to the presence of a rapidly remodeling region that maintains tissue compliance and limits microdamage initiation. Copyright (C) 1999 Elsevier Science Inc.",
keywords = "Bone, Endosseous implants, Fatigue, Histomorphometry, Microdamage, Remodeling",
author = "Huja, {S. S.} and Katona, {T. R.} and Burr, {D. B.} and Garetto, {L. P.} and Roberts, {W. E.}",
year = "1999",
month = "8",
doi = "10.1016/S8756-3282(99)00151-9",
language = "English (US)",
volume = "25",
pages = "217--222",
journal = "Bone",
issn = "8756-3282",
publisher = "Elsevier Inc.",
number = "2",

}

TY - JOUR

T1 - Microdamage adjacent to endosseous implants

AU - Huja, S. S.

AU - Katona, T. R.

AU - Burr, D. B.

AU - Garetto, L. P.

AU - Roberts, W. E.

PY - 1999/8

Y1 - 1999/8

N2 - Intense remodeling occurs in lamellar bone adjacent to osseointegrated endosseous implants. The purpose of this study was to compare microdamage accumulation subsequent to ex vivo fatigue loading of bone that surrounds an endosseous implant, (a) immediately after placement (nonadapted bone) and (b) following a 12 week healing period after placement (adapted bone). We hypothesize that there is less microdamage in the more compliant adapted bone than in the older nonadapted bone. Nonthreaded titanium plasma sprayed (TPS)-coated endosseous implants were placed into dog midfemoral diaphyses and allowed to heal for 12 weeks. Block sections of bone, each containing one implant, were cut anteroposteriorly, resulting in an implant containing lateral cortex, and a medial cortex that was used for testing the nonadapted specimens. Control specimens (n = 14 each for adapted and nonadapted) were loaded at 0 N. Experimental specimens (n = 13, adapted; n = 14, nonadapted) were loaded at 100 N in cantilever bending for 150,000 cycles at 2 Hz, at 37°C on a Bionix 858 testing machine. Specimens were bulk stained with basic fuchsin and 120-140 μm sections were obtained. Crack numerical density (Cr.Dn = Cr.N/B.Ar, #/mm2), crack surface density (Cr.S.Dn = Tt.Cr.Le/B.Ar, mm/mm2), and percent damage area (Dm.Ar = Cr.Ar x 100/B.Ar, mm2/mm2) were measured at x250. Statistically significant differences (p < 0.0001) were seen for Cr.Dn, Cr.S.Dn, and Dm.Ar on the compressed cortices suggesting that adapted bone near the implant accumulated significantly less microdamage than nonadapted bone. Also, the adapted nonloaded control specimens had approximately 20-fold less damage than the respective nonadapted specimens. This study suggests that the compliant adapted bone adjacent to endosseous implants is relatively resistant to fatigue loads. The high success rates of endosseous implants may be due to the presence of a rapidly remodeling region that maintains tissue compliance and limits microdamage initiation. Copyright (C) 1999 Elsevier Science Inc.

AB - Intense remodeling occurs in lamellar bone adjacent to osseointegrated endosseous implants. The purpose of this study was to compare microdamage accumulation subsequent to ex vivo fatigue loading of bone that surrounds an endosseous implant, (a) immediately after placement (nonadapted bone) and (b) following a 12 week healing period after placement (adapted bone). We hypothesize that there is less microdamage in the more compliant adapted bone than in the older nonadapted bone. Nonthreaded titanium plasma sprayed (TPS)-coated endosseous implants were placed into dog midfemoral diaphyses and allowed to heal for 12 weeks. Block sections of bone, each containing one implant, were cut anteroposteriorly, resulting in an implant containing lateral cortex, and a medial cortex that was used for testing the nonadapted specimens. Control specimens (n = 14 each for adapted and nonadapted) were loaded at 0 N. Experimental specimens (n = 13, adapted; n = 14, nonadapted) were loaded at 100 N in cantilever bending for 150,000 cycles at 2 Hz, at 37°C on a Bionix 858 testing machine. Specimens were bulk stained with basic fuchsin and 120-140 μm sections were obtained. Crack numerical density (Cr.Dn = Cr.N/B.Ar, #/mm2), crack surface density (Cr.S.Dn = Tt.Cr.Le/B.Ar, mm/mm2), and percent damage area (Dm.Ar = Cr.Ar x 100/B.Ar, mm2/mm2) were measured at x250. Statistically significant differences (p < 0.0001) were seen for Cr.Dn, Cr.S.Dn, and Dm.Ar on the compressed cortices suggesting that adapted bone near the implant accumulated significantly less microdamage than nonadapted bone. Also, the adapted nonloaded control specimens had approximately 20-fold less damage than the respective nonadapted specimens. This study suggests that the compliant adapted bone adjacent to endosseous implants is relatively resistant to fatigue loads. The high success rates of endosseous implants may be due to the presence of a rapidly remodeling region that maintains tissue compliance and limits microdamage initiation. Copyright (C) 1999 Elsevier Science Inc.

KW - Bone

KW - Endosseous implants

KW - Fatigue

KW - Histomorphometry

KW - Microdamage

KW - Remodeling

UR - http://www.scopus.com/inward/record.url?scp=0033053640&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0033053640&partnerID=8YFLogxK

U2 - 10.1016/S8756-3282(99)00151-9

DO - 10.1016/S8756-3282(99)00151-9

M3 - Article

C2 - 10456388

AN - SCOPUS:0033053640

VL - 25

SP - 217

EP - 222

JO - Bone

JF - Bone

SN - 8756-3282

IS - 2

ER -