Monitoring femoral component installation using vibration testing

Seana Giardini, Phillip Cornwell, R. Meneghini

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

With emerging minimally invasive surgical techniques in total hip arthroplasty, there has been anecdotal evidence of an increase in fractures associated with the insertion of the prosthesis into the femur. The diminished visibility associated with minimally invasive surgical techniques necessitates a greater emphasis on the surgeon's tactile and auditory senses. These senses are used to ascertain the femoral component position of maximum stability and interference fit, as well as to prevent further component impaction and subsequent fracture of the femur. The work described herein attempts to identify a means to supplement the surgeon's tactile and auditory senses by using damage identification techniques normally used in civil and mechanical structures to monitor the insertion process of the prosthesis. It is hypothesized that vibration characteristics of the impact process may be used intra-operatively to determine at what position the femoral component has reached appropriate interference fit and stability in the femur. Such information may be used to prevent further impaction of the femoral component past a threshold that could result in a periprosthetic fracture. A piezoelectric accelerometer and impact hammer will be used to monitor the impact process. The acceleration time history data were analyzed by using low and high pass filters to allow frequency analysis of the time history signals. This paper will summarize features derived from the measured data that will be used to develop an insertion process termination indicator.

Original languageEnglish (US)
Pages (from-to)13-18
Number of pages6
JournalBiomedical Sciences Instrumentation
Volume41
StatePublished - 2005

Fingerprint

Vibration
Thigh
Femur
Touch
Prostheses and Implants
Monitoring
Testing
Periprosthetic Fractures
High pass filters
Arthroplasty
Hammers
Low pass filters
Accelerometers
Visibility
Hip
Surgeons

Keywords

  • Cementless Total Hip Arthroplasty
  • Insertion Monitoring
  • Vibration

ASJC Scopus subject areas

  • Hardware and Architecture

Cite this

Monitoring femoral component installation using vibration testing. / Giardini, Seana; Cornwell, Phillip; Meneghini, R.

In: Biomedical Sciences Instrumentation, Vol. 41, 2005, p. 13-18.

Research output: Contribution to journalArticle

@article{e58ac667d7f2428bbdbfa74aa8f4ad0a,
title = "Monitoring femoral component installation using vibration testing",
abstract = "With emerging minimally invasive surgical techniques in total hip arthroplasty, there has been anecdotal evidence of an increase in fractures associated with the insertion of the prosthesis into the femur. The diminished visibility associated with minimally invasive surgical techniques necessitates a greater emphasis on the surgeon's tactile and auditory senses. These senses are used to ascertain the femoral component position of maximum stability and interference fit, as well as to prevent further component impaction and subsequent fracture of the femur. The work described herein attempts to identify a means to supplement the surgeon's tactile and auditory senses by using damage identification techniques normally used in civil and mechanical structures to monitor the insertion process of the prosthesis. It is hypothesized that vibration characteristics of the impact process may be used intra-operatively to determine at what position the femoral component has reached appropriate interference fit and stability in the femur. Such information may be used to prevent further impaction of the femoral component past a threshold that could result in a periprosthetic fracture. A piezoelectric accelerometer and impact hammer will be used to monitor the impact process. The acceleration time history data were analyzed by using low and high pass filters to allow frequency analysis of the time history signals. This paper will summarize features derived from the measured data that will be used to develop an insertion process termination indicator.",
keywords = "Cementless Total Hip Arthroplasty, Insertion Monitoring, Vibration",
author = "Seana Giardini and Phillip Cornwell and R. Meneghini",
year = "2005",
language = "English (US)",
volume = "41",
pages = "13--18",
journal = "Biomedical Sciences Instrumentation",
issn = "0067-8856",
publisher = "ISA - Instrumentation, Systems, and Automation Society",

}

TY - JOUR

T1 - Monitoring femoral component installation using vibration testing

AU - Giardini, Seana

AU - Cornwell, Phillip

AU - Meneghini, R.

PY - 2005

Y1 - 2005

N2 - With emerging minimally invasive surgical techniques in total hip arthroplasty, there has been anecdotal evidence of an increase in fractures associated with the insertion of the prosthesis into the femur. The diminished visibility associated with minimally invasive surgical techniques necessitates a greater emphasis on the surgeon's tactile and auditory senses. These senses are used to ascertain the femoral component position of maximum stability and interference fit, as well as to prevent further component impaction and subsequent fracture of the femur. The work described herein attempts to identify a means to supplement the surgeon's tactile and auditory senses by using damage identification techniques normally used in civil and mechanical structures to monitor the insertion process of the prosthesis. It is hypothesized that vibration characteristics of the impact process may be used intra-operatively to determine at what position the femoral component has reached appropriate interference fit and stability in the femur. Such information may be used to prevent further impaction of the femoral component past a threshold that could result in a periprosthetic fracture. A piezoelectric accelerometer and impact hammer will be used to monitor the impact process. The acceleration time history data were analyzed by using low and high pass filters to allow frequency analysis of the time history signals. This paper will summarize features derived from the measured data that will be used to develop an insertion process termination indicator.

AB - With emerging minimally invasive surgical techniques in total hip arthroplasty, there has been anecdotal evidence of an increase in fractures associated with the insertion of the prosthesis into the femur. The diminished visibility associated with minimally invasive surgical techniques necessitates a greater emphasis on the surgeon's tactile and auditory senses. These senses are used to ascertain the femoral component position of maximum stability and interference fit, as well as to prevent further component impaction and subsequent fracture of the femur. The work described herein attempts to identify a means to supplement the surgeon's tactile and auditory senses by using damage identification techniques normally used in civil and mechanical structures to monitor the insertion process of the prosthesis. It is hypothesized that vibration characteristics of the impact process may be used intra-operatively to determine at what position the femoral component has reached appropriate interference fit and stability in the femur. Such information may be used to prevent further impaction of the femoral component past a threshold that could result in a periprosthetic fracture. A piezoelectric accelerometer and impact hammer will be used to monitor the impact process. The acceleration time history data were analyzed by using low and high pass filters to allow frequency analysis of the time history signals. This paper will summarize features derived from the measured data that will be used to develop an insertion process termination indicator.

KW - Cementless Total Hip Arthroplasty

KW - Insertion Monitoring

KW - Vibration

UR - http://www.scopus.com/inward/record.url?scp=17044378090&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=17044378090&partnerID=8YFLogxK

M3 - Article

VL - 41

SP - 13

EP - 18

JO - Biomedical Sciences Instrumentation

JF - Biomedical Sciences Instrumentation

SN - 0067-8856