Optimization of hybrid iterative reconstruction level in pediatric body CT

Boaz Karmazyn, Yun Liang, Huisi Ai, George J. Eckert, Mervyn D. Cohen, Matthew Wanner, S. Gregory Jennings

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

OBJECTIVE. The objective of our study was to attempt to optimize the level of hybrid iterative reconstruction (HIR) in pediatric body CT. MATERIALS AND METHODS. One hundred consecutive chest or abdominal CT examinations were selected. For each examination, six series were obtained: one filtered back projection (FBP) and five HIR series (iDose4) levels 2-6. Two pediatric radiologists, blinded to noise measurements, independently chose the optimal HIR level and then rated series quality. We measured CT number (mean in Hounsfield units) and noise (SD in Hounsfield units) changes by placing regions of interest in the liver, muscles, subcutaneous fat, and aorta. A mixed-model analysis-of-variance test was used to analyze correlation of noise reduction with the optimal HIR level compared with baseline FBP noise. RESULTS. One hundred CT examinations were performed of 88 patients (52 females and 36 males) with a mean age of 8.5 years (range, 19 days-18 years); 12 patients had both chest and abdominal CT studies. Radiologists agreed to within one level of HIR in 92 of 100 studies. The mean quality rating was significantly higher for HIR than FBP (3.6 vs 3.3, respectively; p < 0.01). HIR caused minimal (0-0.2%) change in CT numbers. Noise reduction varied among structures and patients. Liver noise reduction positively correlated with baseline noise when the optimal HIR level was used ( p < 0.01). HIR levels were significantly correlated with body weight and effective diameter of the upper abdomen ( p < 0.01). CONCLUSION. HIR, such as iDose4, improves the quality of body CT scans of pediatric patients by decreasing noise; HIR level 3 or 4 is optimal for most studies. The optimal HIR level was less effective in reducing liver noise in children with lower baseline noise.

Original languageEnglish
Pages (from-to)426-431
Number of pages6
JournalAmerican Journal of Roentgenology
Volume202
Issue number2
DOIs
StatePublished - Feb 2014

Fingerprint

Noise
Pediatrics
Liver
Thorax
Subcutaneous Fat
Abdomen
Aorta
Analysis of Variance
Body Weight
Muscles

Keywords

  • ALARA
  • CT
  • Iterative reconstruction
  • Noise level
  • Pediatric imaging

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Optimization of hybrid iterative reconstruction level in pediatric body CT. / Karmazyn, Boaz; Liang, Yun; Ai, Huisi; Eckert, George J.; Cohen, Mervyn D.; Wanner, Matthew; Jennings, S. Gregory.

In: American Journal of Roentgenology, Vol. 202, No. 2, 02.2014, p. 426-431.

Research output: Contribution to journalArticle

Karmazyn, Boaz ; Liang, Yun ; Ai, Huisi ; Eckert, George J. ; Cohen, Mervyn D. ; Wanner, Matthew ; Jennings, S. Gregory. / Optimization of hybrid iterative reconstruction level in pediatric body CT. In: American Journal of Roentgenology. 2014 ; Vol. 202, No. 2. pp. 426-431.
@article{770dd705a27f4fda966346e664ba39c6,
title = "Optimization of hybrid iterative reconstruction level in pediatric body CT",
abstract = "OBJECTIVE. The objective of our study was to attempt to optimize the level of hybrid iterative reconstruction (HIR) in pediatric body CT. MATERIALS AND METHODS. One hundred consecutive chest or abdominal CT examinations were selected. For each examination, six series were obtained: one filtered back projection (FBP) and five HIR series (iDose4) levels 2-6. Two pediatric radiologists, blinded to noise measurements, independently chose the optimal HIR level and then rated series quality. We measured CT number (mean in Hounsfield units) and noise (SD in Hounsfield units) changes by placing regions of interest in the liver, muscles, subcutaneous fat, and aorta. A mixed-model analysis-of-variance test was used to analyze correlation of noise reduction with the optimal HIR level compared with baseline FBP noise. RESULTS. One hundred CT examinations were performed of 88 patients (52 females and 36 males) with a mean age of 8.5 years (range, 19 days-18 years); 12 patients had both chest and abdominal CT studies. Radiologists agreed to within one level of HIR in 92 of 100 studies. The mean quality rating was significantly higher for HIR than FBP (3.6 vs 3.3, respectively; p < 0.01). HIR caused minimal (0-0.2{\%}) change in CT numbers. Noise reduction varied among structures and patients. Liver noise reduction positively correlated with baseline noise when the optimal HIR level was used ( p < 0.01). HIR levels were significantly correlated with body weight and effective diameter of the upper abdomen ( p < 0.01). CONCLUSION. HIR, such as iDose4, improves the quality of body CT scans of pediatric patients by decreasing noise; HIR level 3 or 4 is optimal for most studies. The optimal HIR level was less effective in reducing liver noise in children with lower baseline noise.",
keywords = "ALARA, CT, Iterative reconstruction, Noise level, Pediatric imaging",
author = "Boaz Karmazyn and Yun Liang and Huisi Ai and Eckert, {George J.} and Cohen, {Mervyn D.} and Matthew Wanner and Jennings, {S. Gregory}",
year = "2014",
month = "2",
doi = "10.2214/AJR.13.10721",
language = "English",
volume = "202",
pages = "426--431",
journal = "American Journal of Roentgenology",
issn = "0361-803X",
publisher = "American Roentgen Ray Society",
number = "2",

}

TY - JOUR

T1 - Optimization of hybrid iterative reconstruction level in pediatric body CT

AU - Karmazyn, Boaz

AU - Liang, Yun

AU - Ai, Huisi

AU - Eckert, George J.

AU - Cohen, Mervyn D.

AU - Wanner, Matthew

AU - Jennings, S. Gregory

PY - 2014/2

Y1 - 2014/2

N2 - OBJECTIVE. The objective of our study was to attempt to optimize the level of hybrid iterative reconstruction (HIR) in pediatric body CT. MATERIALS AND METHODS. One hundred consecutive chest or abdominal CT examinations were selected. For each examination, six series were obtained: one filtered back projection (FBP) and five HIR series (iDose4) levels 2-6. Two pediatric radiologists, blinded to noise measurements, independently chose the optimal HIR level and then rated series quality. We measured CT number (mean in Hounsfield units) and noise (SD in Hounsfield units) changes by placing regions of interest in the liver, muscles, subcutaneous fat, and aorta. A mixed-model analysis-of-variance test was used to analyze correlation of noise reduction with the optimal HIR level compared with baseline FBP noise. RESULTS. One hundred CT examinations were performed of 88 patients (52 females and 36 males) with a mean age of 8.5 years (range, 19 days-18 years); 12 patients had both chest and abdominal CT studies. Radiologists agreed to within one level of HIR in 92 of 100 studies. The mean quality rating was significantly higher for HIR than FBP (3.6 vs 3.3, respectively; p < 0.01). HIR caused minimal (0-0.2%) change in CT numbers. Noise reduction varied among structures and patients. Liver noise reduction positively correlated with baseline noise when the optimal HIR level was used ( p < 0.01). HIR levels were significantly correlated with body weight and effective diameter of the upper abdomen ( p < 0.01). CONCLUSION. HIR, such as iDose4, improves the quality of body CT scans of pediatric patients by decreasing noise; HIR level 3 or 4 is optimal for most studies. The optimal HIR level was less effective in reducing liver noise in children with lower baseline noise.

AB - OBJECTIVE. The objective of our study was to attempt to optimize the level of hybrid iterative reconstruction (HIR) in pediatric body CT. MATERIALS AND METHODS. One hundred consecutive chest or abdominal CT examinations were selected. For each examination, six series were obtained: one filtered back projection (FBP) and five HIR series (iDose4) levels 2-6. Two pediatric radiologists, blinded to noise measurements, independently chose the optimal HIR level and then rated series quality. We measured CT number (mean in Hounsfield units) and noise (SD in Hounsfield units) changes by placing regions of interest in the liver, muscles, subcutaneous fat, and aorta. A mixed-model analysis-of-variance test was used to analyze correlation of noise reduction with the optimal HIR level compared with baseline FBP noise. RESULTS. One hundred CT examinations were performed of 88 patients (52 females and 36 males) with a mean age of 8.5 years (range, 19 days-18 years); 12 patients had both chest and abdominal CT studies. Radiologists agreed to within one level of HIR in 92 of 100 studies. The mean quality rating was significantly higher for HIR than FBP (3.6 vs 3.3, respectively; p < 0.01). HIR caused minimal (0-0.2%) change in CT numbers. Noise reduction varied among structures and patients. Liver noise reduction positively correlated with baseline noise when the optimal HIR level was used ( p < 0.01). HIR levels were significantly correlated with body weight and effective diameter of the upper abdomen ( p < 0.01). CONCLUSION. HIR, such as iDose4, improves the quality of body CT scans of pediatric patients by decreasing noise; HIR level 3 or 4 is optimal for most studies. The optimal HIR level was less effective in reducing liver noise in children with lower baseline noise.

KW - ALARA

KW - CT

KW - Iterative reconstruction

KW - Noise level

KW - Pediatric imaging

UR - http://www.scopus.com/inward/record.url?scp=84892746588&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84892746588&partnerID=8YFLogxK

U2 - 10.2214/AJR.13.10721

DO - 10.2214/AJR.13.10721

M3 - Article

C2 - 24450687

AN - SCOPUS:84892746588

VL - 202

SP - 426

EP - 431

JO - American Journal of Roentgenology

JF - American Journal of Roentgenology

SN - 0361-803X

IS - 2

ER -