Use of dMLC for implementation of dynamic respiratory-gated radiation therapy

Eric W. Pepin, Huanmei Wu, Hiroki Shirato

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Purpose: To simulate and evaluate the use of dynamic multileaf collimators (dMLC) in respiratory gating to compensate for baseline drift. Methods: Tumor motion tracking data from 30 lung tumors over 322 treatment fractions was analyzed with the finite state model. A dynamic respiratory gating window was established in real-time by determining the average positions during the previous two end-of-expiration breathing phases and centering the dMLC aperture on a weighted average of these positions. A simulated dMLC with physical motion constraints was used in dynamic gating treatment simulations. Fluence maps were created to provide a statistical description of radiation delivery for each fraction. Duty cycle was also calculated for each fraction. Results: The average duty cycle was 2.3% greater under dynamic gating conditions. Dynamic gating also showed higher fluences and less tumor obstruction. Additionally, dynamic gating required fewer beam toggles and each delivery period was longer on average than with static gating. Conclusions: The use of dynamic gating showed better performance than static gating and the physical constraints of a dMLC were shown to not be an impediment to dynamic gating.

Original languageEnglish
Article number101708
JournalMedical Physics
Volume40
Issue number10
DOIs
StatePublished - 2013

Fingerprint

Radiotherapy
Neoplasms
Respiration
Radiation
Lung
Therapeutics

Keywords

  • Baseline shift
  • Lung cancer
  • Respiratory gating
  • Respiratory motion

ASJC Scopus subject areas

  • Biophysics
  • Radiology Nuclear Medicine and imaging
  • Medicine(all)

Cite this

Use of dMLC for implementation of dynamic respiratory-gated radiation therapy. / Pepin, Eric W.; Wu, Huanmei; Shirato, Hiroki.

In: Medical Physics, Vol. 40, No. 10, 101708, 2013.

Research output: Contribution to journalArticle

@article{5b56ec96e23e416a95ea2cd04c44d73f,
title = "Use of dMLC for implementation of dynamic respiratory-gated radiation therapy",
abstract = "Purpose: To simulate and evaluate the use of dynamic multileaf collimators (dMLC) in respiratory gating to compensate for baseline drift. Methods: Tumor motion tracking data from 30 lung tumors over 322 treatment fractions was analyzed with the finite state model. A dynamic respiratory gating window was established in real-time by determining the average positions during the previous two end-of-expiration breathing phases and centering the dMLC aperture on a weighted average of these positions. A simulated dMLC with physical motion constraints was used in dynamic gating treatment simulations. Fluence maps were created to provide a statistical description of radiation delivery for each fraction. Duty cycle was also calculated for each fraction. Results: The average duty cycle was 2.3{\%} greater under dynamic gating conditions. Dynamic gating also showed higher fluences and less tumor obstruction. Additionally, dynamic gating required fewer beam toggles and each delivery period was longer on average than with static gating. Conclusions: The use of dynamic gating showed better performance than static gating and the physical constraints of a dMLC were shown to not be an impediment to dynamic gating.",
keywords = "Baseline shift, Lung cancer, Respiratory gating, Respiratory motion",
author = "Pepin, {Eric W.} and Huanmei Wu and Hiroki Shirato",
year = "2013",
doi = "10.1118/1.4820534",
language = "English",
volume = "40",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "10",

}

TY - JOUR

T1 - Use of dMLC for implementation of dynamic respiratory-gated radiation therapy

AU - Pepin, Eric W.

AU - Wu, Huanmei

AU - Shirato, Hiroki

PY - 2013

Y1 - 2013

N2 - Purpose: To simulate and evaluate the use of dynamic multileaf collimators (dMLC) in respiratory gating to compensate for baseline drift. Methods: Tumor motion tracking data from 30 lung tumors over 322 treatment fractions was analyzed with the finite state model. A dynamic respiratory gating window was established in real-time by determining the average positions during the previous two end-of-expiration breathing phases and centering the dMLC aperture on a weighted average of these positions. A simulated dMLC with physical motion constraints was used in dynamic gating treatment simulations. Fluence maps were created to provide a statistical description of radiation delivery for each fraction. Duty cycle was also calculated for each fraction. Results: The average duty cycle was 2.3% greater under dynamic gating conditions. Dynamic gating also showed higher fluences and less tumor obstruction. Additionally, dynamic gating required fewer beam toggles and each delivery period was longer on average than with static gating. Conclusions: The use of dynamic gating showed better performance than static gating and the physical constraints of a dMLC were shown to not be an impediment to dynamic gating.

AB - Purpose: To simulate and evaluate the use of dynamic multileaf collimators (dMLC) in respiratory gating to compensate for baseline drift. Methods: Tumor motion tracking data from 30 lung tumors over 322 treatment fractions was analyzed with the finite state model. A dynamic respiratory gating window was established in real-time by determining the average positions during the previous two end-of-expiration breathing phases and centering the dMLC aperture on a weighted average of these positions. A simulated dMLC with physical motion constraints was used in dynamic gating treatment simulations. Fluence maps were created to provide a statistical description of radiation delivery for each fraction. Duty cycle was also calculated for each fraction. Results: The average duty cycle was 2.3% greater under dynamic gating conditions. Dynamic gating also showed higher fluences and less tumor obstruction. Additionally, dynamic gating required fewer beam toggles and each delivery period was longer on average than with static gating. Conclusions: The use of dynamic gating showed better performance than static gating and the physical constraints of a dMLC were shown to not be an impediment to dynamic gating.

KW - Baseline shift

KW - Lung cancer

KW - Respiratory gating

KW - Respiratory motion

UR - http://www.scopus.com/inward/record.url?scp=84885749150&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84885749150&partnerID=8YFLogxK

U2 - 10.1118/1.4820534

DO - 10.1118/1.4820534

M3 - Article

VL - 40

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 10

M1 - 101708

ER -